高交联聚乙烯相比传统聚乙烯在非骨水泥全髋关节置换术中平均12年随访降低骨溶解发生率和磨损相关再手术率
该回顾性队列研究比较了非骨水泥全髋关节置换术中使用高交联聚乙烯(HXLPE)与传统聚乙烯(CPE)内衬的患者,平均随访12年。结果显示HXLPE组磨损相关再手术生存率100% vs CPE组91.4%(P=0.007),骨溶解生存率100% vs 36.2%(P<0.001),HXLPE组磨损率显著降低(0.035 mm/y vs 0.118 mm/y)。
要点
- 平均12年随访,HXLPE组磨损相关再手术生存率100%,CPE组91.4%(P=0.007)。
- HXLPE组骨溶解生存率100%,CPE组36.2%(P<0.001)。
- HXLPE组磨损率0.035 mm/y,CPE组0.118 mm/y,HXLPE组显著降低。
核心结论
- 使用同一制造商的高交联聚乙烯(HXLPE)内衬在平均12年随访中展现出降低磨损相关再手术率的临床优势。
研究设计与样本
- 回顾性队列研究,纳入45-70岁使用26 mm直径钴铬球头的非骨水泥全髋关节置换术患者,共67例患者(79髋;HXLPE组41髋,CPE组38髋),至少随访10年,平均随访12年。
主要发现
- 以磨损相关再手术为终点,HXLPE组平均12年生存率100%,CPE组91.4%(P=0.007)。
- 以骨溶解为终点,HXLPE组平均12年生存率100%,CPE组36.2%(P<0.001)。
- HXLPE组磨损率0.035 mm/y,CPE组0.118 mm/y,HXLPE组显著降低。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: A number of studies on total hip arthroplasty have compared highly cross-linked polyethylene (HXLPE) with conventional polyethylene (CPE) liners beyond 10 years. However, the impact of HXLPE on the wear-related reoperation rate is unclear. The purpose of this study was to evaluate the clinical advantage of using a single manufacturer's HXLPE in terms of reducing the reoperation rate. METHODS: The study was a follow-up retrospective cohort study over a mean of 12 years that examined patients aged 45-70 years with cementless total hip arthroplasty using a 26-mm-diameter cobalt-chromium head. Sixty-seven patients (79 hips; HXLPE group = 41 hips, CPE group = 38 hips) were evaluated for a minimum 10-year follow-up. Kaplan-Meier survival analysis was performed, with wear-related reoperations and radiographic osteolysis serving as the end points. The polyethylene wear rate was also assessed. RESULTS: The mean 12-year follow-up rates of survivorship that were evaluated using wear-related reoperations as the end point were 100% and 91.4% in the HXLPE and CPE groups, respectively (P = .007), and the mean 12-year follow-up rates of survivorship with osteolysis as the end point were 100% and 36.2%, respectively (P < .001). Compared with the CPE group, the HXLPE group presented a significantly reduced wear rate (HXLPE group, 0.035 mm/y; CPE group, 0.118 mm/y). CONCLUSION: A unique strength of this study is that we assessed a single manufacturer's HXLPE while keeping most other implant parameters uniform. This study reveals the clinical advantage of using a single manufacturer's HXLPE in terms of a reduced wear-related reoperation rate at a mean 12-year follow-up.
引用
- Tsukamoto M, Mori T, Ohnishi H, et al. Highly Cross-Linked Polyethylene Reduces Osteolysis Incidence and Wear-Related Reoperation Rate in Cementless Total Hip Arthroplasty Compared With Conventional Polyethylene at a Mean 12-Year Follow-Up. J Arthroplasty. 2017. PMID: 28734615. doi:10.1016/j.arth.2017.06.047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。