取出的金属对金属髋关节假体的微结构与表面损伤
对来自多中心取出项目的20例金属对金属髋关节假体的髋臼与股骨部件进行多模式显微镜检查,在每种关节部件的25个区域观察金相与加工特征。结果显示气孔、缩孔、碳化物脱落孔洞、表面碳化物、嵌入颗粒、划痕、磨损痕迹、表面变色、表面沉积物及摩擦化学反应层广泛分布于所有植入部件的大部分承载面。髋臼杯与球头表面空洞及杯划痕与配对间隙显著相关,低间隙假体的杯与球头表面损伤更重;表面髋关节置换与全髋关节置换之间或不同配对直径之间,上述缺陷的发生率和严重程度无其他显著相关或差异。
要点
- 所有取出的金属对金属髋关节假体承载面均广泛存在气孔、缩孔、碳化物脱落孔洞、表面碳化物、嵌入颗粒、划痕、磨损痕迹、表面变色、表面沉积物及摩擦化学反应层。
- 髋臼杯与球头表面空洞及杯划痕与配对间隙显著相关,低间隙假体的杯与球头表面损伤更重。
- 表面髋关节置换与全髋关节置换之间,或不同配对直径之间,上述缺陷的发生率和严重程度无其他显著相关或差异。
核心结论
- 取出的金属对金属髋关节假体中,部分金相特征和表面损伤显著存在;作者认为希望改善这些植入物的结构和金相特征以避免这些效应并优化其性能。
研究设计与样本
- 从多中心取出项目收集20例金属对金属髋关节假体的髋臼和股骨部件,用肉眼、共聚焦显微镜、垂直扫描干涉仪、扫描电子显微镜、背散射电子成像和能量色散X射线光谱检查,每种关节部件检查25个区域。
主要发现
- 气孔、缩孔、碳化物脱落孔洞、表面碳化物、嵌入颗粒、划痕、磨损痕迹、表面变色、表面沉积物和摩擦化学反应层广泛分布于所有植入部件的大部分承载面。
- 髋臼杯与球头表面空洞及杯划痕与配对间隙显著相关。
- 低间隙假体的髋臼杯和球头表面损伤更重。
- 表面髋关节置换与全髋关节置换之间,或不同配对直径之间,上述缺陷的发生率和严重程度无其他显著相关或差异。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Besides promising results of metal-on-metal (MOM) hip arthroplasty (HA), frequent failures have been reported even in the short term. Many host, surgical, design, metallurgical, and processing factors have been evoked in the base of these events. We have tried to characterize and to evaluate metallurgical and processing features present in this type of implants. METHODS: The acetabular and femoral components of 20 MOM HAs collected from a multicenter retrieval program were examined. All the specimens were inspected with naked eye, with confocal microscopy and vertical scanning interferometry, scanning electron microscopy, back-scattered electron imaging, and energy-dispersive X-ray spectroscopy, in 25 zones of each articular component. RESULTS: Gas pores, shrinkage voids and holes of detached carbides, carbides on surface, embedded particles, scratches and marks of wear, surface discoloration, surface deposits, and tribochemical reaction layers were widely dispersed through a substantial percentage of the total bearing surface in all the implanted components. Surface cup and head voids, and cup scratches showed significant correlation with the clearance of pair. A higher surface damage of the cup and head was observed mainly in the low clearance prostheses. There was no other significant correlation or difference in the incidence and importance of any of these defects between resurfacing hip arthroplasties and total hip arthroplasties, or according to the pair diameter. CONCLUSION: Some metallurgical features and surface damage were significantly present in the retrieved implants of MoM HAs. It would be desirable to improve the structure and metallurgical characteristics of these implants to avoid those effects and optimize their performance.
引用
- Fernández-Fairen M, Punset M, Murcia-Asensio A, et al. Microstructure and Surface Damage in Retrieved Metal-on-Metal Hip Arthroplasties. J Arthroplasty. 2017. PMID: 28754580. doi:10.1016/j.arth.2017.06.044