陶瓷头锥形适配套筒的微动与腐蚀损伤:一项取回研究

Fretting and Corrosion Damage in Taper Adapter Sleeves for Ceramic Heads: A Retrieval Study.
PubMed 文献 · J Arthroplasty · 2017-04-26

本研究对37例翻修术中取回的带套筒陶瓷头进行微动腐蚀评估,植入时间0.0-3.3年。结果显示92%内锥、19%外锥和78%股骨柄存在轻度至中度微动腐蚀,仅1例既往保留的股骨柄锥部出现严重腐蚀,套筒均无严重腐蚀。原发与翻修手术使用的套筒腐蚀损伤无差异,作者认为从体内腐蚀角度看套筒是翻修时恢复股骨柄锥形的合理方案。

要点

  • 37例取回带套筒陶瓷头中,92%内锥、19%外锥和78%股骨柄见轻度至中度微动腐蚀(评分2-3)。
  • 仅1例既往翻修保留的股骨柄锥部见严重微动腐蚀,取回套筒均无严重腐蚀。
  • 原发与翻修手术使用的套筒腐蚀损伤无差异。

核心结论

  • 本研究取回套筒陶瓷头的微动腐蚀评分以轻度为主,低于金属对聚乙烯轴承中钴铬头的报道值。
  • 套筒虽拟用于翻修,但短期体内腐蚀行为在原发与翻修手术应用中相似。
  • 从体内腐蚀角度看,套筒是翻修时恢复股骨柄锥形的合理方案。

研究设计与样本

  • 研究纳入37例翻修术中取回的带套筒陶瓷头,股骨头与套筒植入时间0.0-3.3年。
  • 取回原因以不稳定、感染和松动为主,50%的取回物在原发手术中植入。
  • 采用Goldberg-Higgs半定量评分系统评估微动腐蚀。

主要发现

  • 轻度至中度微动腐蚀(评分2-3)见于92%的内锥、19%的外锥和78%的股骨柄。
  • 严重微动腐蚀见于1例既往翻修术中保留的股骨柄锥部,取回套筒均无严重腐蚀。
  • 原发手术与翻修手术使用的套筒在腐蚀损伤上无差异。

关键数值与推荐

  • 微动腐蚀采用Goldberg-Higgs半定量评分系统评估,评分2-3定义为轻度至中度。

临床怎么用

  • 作者认为,从体内腐蚀角度看,套筒是翻修手术中恢复股骨柄锥形的合理方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: During revision surgery with a well-fixed stem, a titanium sleeve can be used in conjunction with a ceramic head to achieve better stress distribution across the taper surface. In vitro testing suggests that corrosion is not a concern in sleeved ceramic heads; however, little is known about the in vivo fretting corrosion of the sleeves. The purpose of this study was to investigate fretting corrosion in sleeved ceramic heads in retrieved total hip arthroplasties. METHODS: Thirty-seven sleeved ceramic heads were collected during revision. The femoral heads and sleeves were implanted 0.0-3.3 years. The implants were revised predominantly for instability, infection, and loosening. Fifty percent of the retrievals were implanted during a primary surgery. Fretting corrosion was assessed using the Goldberg-Higgs semiquantitative scoring system. RESULTS: Mild-to-moderate fretting corrosion scores (score = 2-3) were observed in 92% of internal tapers, 19% of external tapers, and 78% of the stems. Severe fretting corrosion was observed in 1 stem trunnion that was previously retained during revision surgery and none of the retrieved sleeves. There was no difference in corrosion damage of sleeves used in primary or revision surgery. CONCLUSION: The fretting corrosion scores in this study were predominantly mild and lower than reported fretting scores of cobalt-chrome heads in metal-on-polyethylene bearings. Although intended for use in revisions, we found that the short-term in vivo corrosion behavior of the sleeves was similar in both primary and revision surgery applications. From an in vivo corrosion perspective, sleeves are a reasonable solution for restoring the stem taper during revision surgery.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. MacDonald DW, Chen AF, Lee GC, et al. Fretting and Corrosion Damage in Taper Adapter Sleeves for Ceramic Heads: A Retrieval Study. J Arthroplasty. 2017. PMID: 28668212. doi:10.1016/j.arth.2017.04.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。