三种不同全膝关节置换假体设计的滑车表面损伤与磨损差异

Differences in Trochlear Surface Damage and Wear Between Three Different Total Knee Arthroplasty Designs.
PubMed 文献 · J Arthroplasty · 2017-07-05

本研究对18例取出的股骨部件(Triathlon、SIGMA和GENESIS II三种现代设计)按体内时间、年龄、性别和体重指数匹配,采用目视检查、光学显微镜和光学轮廓术评估滑车磨损与表面损伤。目视检查各组表面损伤无显著差异;光学轮廓术显示取出的Triathlon(P=0.002)和SIGMA(P=0.009)部件较相应参照部件显著更粗糙,GENESIS II取出部件与参照部件无显著差异(P=0.48),且SIGMA取出部件的粗糙度显著高于Triathlon或GENESIS II部件。

要点

  • 目视检查显示所有假体类型均有表面损伤证据,但各组间表面损伤和磨损无显著差异。
  • 光学轮廓术显示取出Triathlon(P=0.002)和SIGMA(P=0.009)部件较参照部件显著更粗糙。
  • 取出SIGMA部件粗糙度显著高于Triathlon或GENESIS II部件,尤其近端滑车沟和内侧滑车。

核心结论

  • 取出的SIGMA部件较Triathlon或GENESIS II部件似乎具有更高的粗糙度。
  • 滑车粗糙度增加的病因学和长期影响需进一步研究。

研究设计与样本

  • 对18例取出的股骨部件(三种现代设计:Triathlon、SIGMA和GENESIS II)按体内时间、年龄、性别和体重指数进行匹配。
  • 采用目视检查、光学显微镜和光学轮廓术评估滑车磨损与表面损伤。

主要发现

  • 目视检查显示所有假体类型均有表面损伤证据,各组间表面损伤和磨损无显著差异。
  • 光学轮廓术显示取出Triathlon部件较相应参照部件显著更粗糙(P=0.002)。
  • 光学轮廓术显示取出SIGMA部件较相应参照部件显著更粗糙(P=0.009)。
  • 取出GENESIS II部件与参照部件无显著粗糙度差异(P=0.48)。
  • 光学轮廓术分析显示取出SIGMA部件粗糙度显著高于Triathlon或GENESIS II部件,尤其在近端滑车沟和内侧滑车。

局限性

  • 滑车粗糙度增加的病因学和长期影响需进一步研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Trochlear design plays a role in patellofemoral kinematics. Little is known regarding differences in the contact properties between modern designs. The purpose of the present study was to analyze patellofemoral joint contact by identifying areas of joint surface damage and wear in retrieved femoral components of 3 modern designs. MATERIALS: Eighteen retrieved femoral components featuring 3 different modern designs (Triathlon, SIGMA, and GENESIS II) were matched based on time-in-vivo, age, gender, and body mass index. Trochlear wear and surface damage were assessed using visual inspection, light microscopy, and light profilometry. RESULTS: Visual inspection of the femoral components showed evidence of surface damage in all implant types. No significant differences between the groups were found with respect to surface damage and wear on visual inspection. Light profilometry of retrieved components showed that retrieved Triathlon (P = .002) and SIGMA (P = .009) components were significantly rougher than the corresponding reference components. The GENESIS II retrieved components were not substantially rougher than the reference components (P = .48). Light profilometry analysis showed significantly increased roughness of retrieved SIGMA components compared with Triathlon or GENESIS II components, particularly in the proximal trochlear groove and the medial trochlea. CONCLUSION: As the volume and patient demands for total knee arthroplasty increase, a greater understanding of the patellofemoral joint is warranted. Patellofemoral kinematics can have an effect on the surface characteristics of total knee arthroplasty components. Retrieved SIGMA components appeared to have increased roughness compared with Triathlon or GENESIS II components. The etiology and long-term effects of increased trochlear roughness require further investigation.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Matz J, Howard JL, Sisko ZW, et al. Differences in Trochlear Surface Damage and Wear Between Three Different Total Knee Arthroplasty Designs. J Arthroplasty. 2017. PMID: 28750858. doi:10.1016/j.arth.2017.06.037