污染与清洁对全髋关节置换术组配式股骨头锥度固定强度的影响
本研究将金属或陶瓷股骨头敲击到钛合金股骨柄锥度上,界面分别经清洁或污染(脂肪或盐水)处理,并比较清洁后干燥与否的拔出力。结果显示脂肪污染显著降低两种股骨头的锥度拔出力,盐水污染则与未污染无显著差异;清洁并干燥后拔出力与未污染相似,而不干燥则拔出力变异大,强调干燥对获得充分且可重复的固定强度至关重要。
要点
- 脂肪污染显著降低金属和陶瓷股骨头的锥度拔出力。
- 盐水污染后的拔出力与未污染锥度无显著差异。
- 清洁并干燥后的拔出力与未污染锥度相似,不干燥则变异大。
核心结论
- 术中锥度界面污染可能难以发现,但对锥度固定强度有重大影响。
- 组装前用盐水清洁股骨柄锥度并用纱布干燥,可使锥度强度达到原始组件的水平。
- 不干燥锥度会导致拔出力变异大,强调干燥对获得充分且可重复的固定强度至关重要。
研究设计与样本
- 体外实验:将金属或陶瓷股骨头敲击到钛合金股骨柄锥度上,界面分为清洁或污染(脂肪或盐水溶液),并在清洁和干燥污染界面后重复相同操作。
- 采用拔出力确定污染和清洁对锥度强度的影响。
主要发现
- 脂肪污染后,金属和陶瓷股骨头的拔出力均显著低于未污染界面。
- 盐水溶液污染后的拔出力与未污染锥度无显著差异。
- 观察到锥度强度变异大;清洁并干燥后的锥度拔出力与未污染锥度相似(两种股骨头材料均如此)。
临床怎么用
- 作者结论指出:组装前直接用盐水清洁股骨柄锥度并用纱布干燥,可使锥度强度达到原始组件的水平。
- 作者结论指出:不干燥锥度会导致拔出力变异大,干燥对获得充分且可重复的固定强度至关重要。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Intraoperative interface contamination of modular head-stem taper junctions of hip implants can lead to poor fixation strength, causing fretting and crevice corrosion or even stem taper fracture. Careful cleaning before assembly should help to reduce these problems. The purpose of this study was to determine the effect of cleaning (with and without drying) contaminated taper interfaces on the taper fixation strength. METHODS: Metal or ceramic heads were impacted onto titanium alloy stem tapers with cleaned or contaminated (fat or saline solution) interfaces. The same procedure was performed after cleaning and drying the contaminated interfaces. Pull-off force was used to determine the influence of contamination and cleaning on the taper strength. RESULTS: Pull-off forces after contamination with fat were significantly lower than those for uncontaminated interfaces for both head materials. Pull-off forces after application of saline solution were not significantly different from those for uncontaminated tapers. However, a large variation in taper strength was observed, pull-off forces for cleaned and dried tapers were similar to those for uncontaminated tapers for both head materials. CONCLUSION: Intraoperative contamination of taper interfaces may be difficult to detect but has a major influence on taper fixation strength. Cleaning of the stem taper with saline solution and drying with gauze directly before assembly allows the taper strength of the pristine components to be achieved. Not drying the taper results in a large variation in pull-off forces, emphasizing that drying is essential for sufficient and reproducible fixation strength.
引用
- Krull A, Morlock MM, Bishop NE, et al. The Influence of Contamination and Cleaning on the Strength of Modular Head Taper Fixation in Total Hip Arthroplasty. J Arthroplasty. 2017. PMID: 28625686. doi:10.1016/j.arth.2017.05.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。