全膝关节置换术中氧化锆与钴合金承重面的比较:对取出的聚乙烯衬垫进行三维激光扫描

Oxidised zirconium versus cobalt alloy bearing surfaces in total knee arthroplasty: 3D laser scanning of retrieved polyethylene inserts.
PubMed 文献 · Bone Joint J · 2017-01-01

研究将20例与氧化锆(OxZr)股骨假体配合的取出胫骨衬垫与20例钴合金(CoCr)假体衬垫按患者年龄、体重指数、植入时长和翻修诊断匹配,用激光扫描比较聚乙烯体积丢失。OxZr组聚乙烯体积丢失122 mm³(标准差87),CoCr组170 mm³(标准差96),差异有统计学意义(p=0.033)。

要点

  • OxZr组聚乙烯体积丢失122 mm³(标准差87),低于CoCr组的170 mm³(标准差96),p=0.033。
  • 内侧间室OxZr组72 mm³(标准差67)vs CoCr组92 mm³(标准差60),p=0.096,差异无统计学意义。
  • 外侧间室OxZr组49 mm³(标准差36)vs CoCr组79 mm³(标准差61),p=0.096,差异无统计学意义。

核心结论

  • 结果与早期体外测试和视觉取出物分析一致,提示OxZr股骨假体的聚乙烯体积丢失较低。
  • 由于OxZr和CoCr均为硬表面,预期产生相近的聚乙烯蠕变,体积丢失差异可能反映这些衬垫体内磨损的差异。

研究设计与样本

  • 将20例与OxZr股骨假体配合的取出胫骨衬垫与20例CoCr配合的衬垫按患者年龄、体重指数、植入时长和翻修诊断进行匹配。
  • 使用激光扫描比较取出衬垫与原始衬垫关节面尺寸的变化,计算并比较两组取出表面与原始表面之间的体积差异。

主要发现

  • OxZr组聚乙烯体积丢失122 mm³(标准差87),CoCr组170 mm³(标准差96),p=0.033。
  • 内侧间室体积丢失OxZr组72 mm³(标准差67)vs CoCr组92 mm³(标准差60),p=0.096。
  • 外侧间室体积丢失OxZr组49 mm³(标准差36)vs CoCr组79 mm³(标准差61),p=0.096。
  • 内侧和外侧间室的组间差异均无统计学意义。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty. MATERIALS AND METHODS: A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared. RESULTS: The loss of polyethylene volume was 122 mm3 (standard deviation (sd) 87) in the OxZr group and 170 mm3 (sd 96) in the CoCr group (p = 0.033). The volume loss in the OxZr group was also lower in the medial (72 mm3 (sd 67) versus 92 mm3 (sd 60); p = 0.096) and lateral (49 mm3 (sd 36) versus 79 mm3 (sd 61); p = 0.096) compartments separately, but these differences were not significant. CONCLUSION: Our results corroborate earlier findings from in vitro testing and visual retrieval analysis which suggest that polyethylene volume loss is lower with OxZr femoral components. Since both OxZr and CoCr are hard surfaces that would be expected to create comparable amounts of polyethylene creep, the differences in volume loss may reflect differences in the in vivo wear of these inserts. Cite this article: Bone Joint J 2017;99-B:793-8.

来源:Bone Joint J,查看原文。版权归原作者所有。