透视引导下髋臼假体定位:能否降低肥胖患者假体位置不良的风险?
该研究回顾性分析了1599例连续接受透视引导下前入路全髋关节置换术患者的术后站立位骨盆正位X线片,按体重指数分为非肥胖、肥胖和重度肥胖三组,比较髋臼假体倾斜角和前倾角。结果显示三组间假体位置无显著差异,透视引导有助于实现目标髋臼假体方位,且不受患者体重指数影响。
要点
- 透视引导下前入路全髋关节置换术后,95.0%的髋臼假体倾斜角位于安全区,95.7%的前倾角位于安全区,91.2%的倾斜角和前倾角均位于安全区。
- 非肥胖、肥胖和重度肥胖组的平均倾斜角分别为37.5°、37.9°和39.9°,平均前倾角分别为16.1°、16.5°和16.0°。
- 体重指数与髋臼假体位置无显著关系:倾斜角P=0.867,前倾角P=0.673,倾斜角和前倾角均P=0.624。
核心结论
- 透视是直接前入路全髋关节置换术中实现目标髋臼假体方位的有用工具,且不受患者体重指数影响。
研究设计与样本
- 回顾性分析前瞻性收集的数据,纳入六年期间连续接受透视引导下前入路全髋关节置换术的1599例患者。
- 根据体重指数将患者分为三组:非肥胖(<30 kg/m²)1065例、肥胖(30-40 kg/m²)506例、重度肥胖(>40 kg/m²)28例。
- 采用Barrack等描述的方法测量术后站立位骨盆正位X线片上的髋臼假体倾斜角和前倾角,术中目标范围倾斜角为30°-50°、前倾角为5°-25°。
主要发现
- 总体平均倾斜角为37.7°,平均前倾角为16.3°。
- 95.0%的髋臼假体倾斜角位于安全区,95.7%的前倾角位于安全区,91.2%的倾斜角和前倾角均位于安全区。
- 非肥胖组、肥胖组和重度肥胖组的平均倾斜角分别为37.5°、37.9°和39.9°,平均前倾角分别为16.1°、16.5°和16.0°。
- 体重指数与髋臼假体位置无显著关系:倾斜角P=0.867,前倾角P=0.673,倾斜角和前倾角均P=0.624。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Obesity has been described as an independent risk factor for acetabular component malpositioning. The purpose of this study was to determine if this could be overcome by use of fluoroscopic navigation in total hip arthroplasty (THA). METHODS: The first, postoperative, standing, anteroposterior pelvis radiographs from 1599 consecutive patients who underwent fluoroscopic-guided THAs via anterior approach during a six-year period were obtained. We retrospectively reviewed this prospectively collected data. Inclination and anteversion were measured as described by Barrack et al. Intraoperative target ranges for inclination and anteversion angles were 30°-50° and 5°-25°. Patients were divided into 3 cohorts by body mass index, nonobese (<30 kg/m2) obese (between 30 and 40 kg/m2) and morbidly obese (>40 kg/m2), looking for significant differences in acetabular component positioning. RESULTS: Of the 1599 patients, 1065 were nonobese, 506 obese, and 28 morbidly obese. Overall, average inclination was 37.7° and anteversion was 16.3°. Regarding inclination, 95.0% of cups were positioned in the safe zone, 95.7% for anteversion, and 91.2% for both inclination and anteversion. Analysis of each cohort individually revealed an average inclination of 37.5° in the nonobese, 37.9° in the obese, and 39.9° in the morbidly obese patients. For anteversion, each group's averages were 16.1°, 16.5°, and 16.0°, respectively. There was no significant relationship between a patient's body mass index and cup position for inclination (P = .867), anteversion (P = .673), or both inclination and anteversion (P = .624). CONCLUSION: Fluoroscopy is a useful tool for achieving a targeted acetabular component orientation in direct anterior THA, irrespective of patient BMI.