BIOLOX delta 陶瓷对陶瓷全髋关节置换术的中期结果
这项前瞻性研究纳入2009年3月至2011年3月接受delta陶瓷全髋关节置换术的274例患者(310髋),评估陶瓷碎裂率、术后噪声及中期结果与生存率。平均随访66.5个月时,Harris髋关节评分从术前47.1分改善至93.8分,6年再手术生存率为99.0%(95%CI 97.8%至100%),delta陶瓷碎裂率为0.3%(286髋中1髋)。
要点
- 平均随访66.5个月,Harris髋关节评分从术前47.1分改善至93.8分。
- 6年以任何原因再手术为终点的生存率为99.0%(95%CI 97.8%至100%)。
- delta陶瓷碎裂率为0.3%(286髋中1髋),因陶瓷内衬安放不当所致。
核心结论
- delta陶瓷碎裂率为0.3%(286髋中1髋)。
- 既往陶瓷对陶瓷全髋关节置换术中以陶瓷头碎裂为主,而delta陶瓷内衬因安放不当导致的碎裂值得关注。
研究设计与样本
- 前瞻性研究,2009年3月至2011年3月共274例患者(310髋)接受非骨水泥型delta陶瓷股骨头和内衬全髋关节置换术。
- 随访时记录临床和影像学结果,采用Kaplan-Meier分析估计生存率。
主要发现
- 4例患者(4髋)死亡,18例患者(20髋)在5年内失访;其余252例患者(286髋)平均随访66.5个月(60至84)。
- 144例男性(166髋)和108例女性(120髋),手术时平均年龄49.7岁(16至83)。
- 平均术前Harris髋关节评分47.1分改善至末次随访93.8分。
- 6例患者报告7髋出现吱吱声,但均不可闻。
- 52髋(18.2%)见累及Gruen 1区和/或7区的放射透亮线;无髋可检测到磨损、局灶性骨溶解或松动迹象。
关键数值与推荐
- 1髋因陶瓷内衬碎裂行翻修,发生于术中陶瓷内衬安放不当未被发现。
- 1髋因股骨假体周围骨折行翻修,1髋因假体周围关节感染接受治疗。
- 6年以任何原因再手术为终点的生存率为99.0%(95%CI 97.8%至100%)。
局限性
- 4例患者(4髋)死亡,18例患者(20髋)在5年内失访。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: We conducted a prospective study of a delta ceramic total hip arthroplasty (THA) to determine the rate of ceramic fracture, to characterise post-operative noise, and to evaluate the mid-term results and survivorship. PATIENTS AND METHODS: Between March 2009 and March 2011, 274 patients (310 hips) underwent cementless THA using a delta ceramic femoral head and liner. At each follow-up, clinical and radiological outcomes were recorded. A Kaplan-Meier analysis was undertaken to estimate survival. RESULTS: Four patients (four hips) died and 18 patients (20 hips) were lost to follow-up within five years. The remaining 252 patients (286 hips) were followed for a mean of 66.5 months (60 to 84). There were 144 men (166 hips) and 108 women (120 hips) with a mean age of 49.7 years (16 to 83) at surgery. The mean pre-operative Harris Hip Score of 47.1 points improved to 93.8 points at final follow-up. Six patients reported squeaking in seven hips; however, none were audible. Radiolucent lines involving Gruen zones one and/or seven were seen in 52 hips (18.2%). No hip had detectable wear, focal osteolysis or signs of loosening. One hip was revised because of fracture of the ceramic liner, which occurred due to an undetected malseating of the ceramic liner at the time of surgery. One hip was revised for a periprosthetic fracture of the femur, and one hip was treated for periprosthetic joint infection. The six-year survivorship with re-operation for any reason as the endpoint was 99.0% (95% confidence interval 97.8% to 100%). DISCUSSION: The rate of delta ceramic fracture was 0.3% (one of 286). While ceramic head fracture was dominant in previous ceramic-on-ceramic THA, fracture of the delta ceramic liner due to malseating is a concern. Cite this article: Bone Joint J 2017;99-B:741-8.