文献速递脊柱外科小儿骨科

早发性脊柱侧凸磁控生长棒手术中非计划再手术:至少两年随访

Unplanned Reoperations in Magnetically Controlled Growing Rod Surgery for Early Onset Scoliosis With a Minimum of Two-Year Follow-Up.
PubMed 文献 · Spine (Phila Pa 1976) · 2017-01-01

这项多中心研究回顾性分析了2009至2012年间30例接受磁控生长棒(MCGR)植入的早发性脊柱侧凸(EOS)患者的前瞻性收集数据,至少随访2年。结果显示14例(46.7%)发生非计划再手术,平均再手术时间为术后23个月,原因包括撑开失败、近端基础失败、断棒和感染;撑开频率更密(1周至2个月)与再手术率更高相关(71% vs. 25%)。

要点

  • 30例EOS患者接受MCGR植入,平均随访37个月,14例(46.7%)发生非计划再手术。
  • 非计划再手术原因包括撑开失败、近端基础失败、断棒和感染。
  • 撑开间隔1周至2个月者再手术率71%,高于3至6个月者的25%。

核心结论

  • MCGR手术可能与非计划再手术相关,更频繁的撑开可能是危险因素。
  • 需要与传统生长棒进行长期对比研究以评估该植入物的有效性。

研究设计与样本

  • 回顾性分析多中心前瞻性收集的临床和影像学数据,至少随访2年。
  • 2009年至2012年间,6家机构共30例EOS患者接受MCGR植入。

主要发现

  • 患者接受MCGR植入的平均年龄为7.2岁,平均随访37个月。
  • 14例(46.7%)在随访期间接受非计划再手术,平均再手术时间为初次手术后23个月(范围5-48个月)。
  • 非计划再手术的原因包括撑开失败、近端基础失败、断棒和感染。
  • 撑开频率在1周至2个月之间的患者再手术率为71%,撑开频率在3至6个月之间的患者为25%。

局限性

  • 需要与传统生长棒进行长期对比研究以评估该植入物的有效性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: A retrospective review of prospectively collected clinical and radiologic data of patients with magnetically controlled growing rods (MCGRs) from a multi-centered study with a minimum of 2-year follow-up. OBJECTIVE: The aim of this study was to describe the incidence and causes of unplanned reoperations and to report the outcomes of patients treated with MCGR for early-onset scoliosis (EOS). SUMMARY OF BACKGROUND DATA: Published clinical studies have demonstrated that MCGR is safe and effective for curvature control of EOS, and can avoid repeated surgeries for distractions. However, there have been no reports on the unplanned reoperations and complications of MCGR for EOS with a large series of patients. METHODS: Between 2009 and 2012, 30 patients with EOS underwent MCGR implantation in six institutions. A retrospective review of prospectively collected clinical and radiologic data with a minimum of 2-year follow-up was conducted. Demographic data, radiologic measurements, unplanned reoperations, and other complications were noted. Risk factors for unplanned reoperations were analyzed. RESULTS: Patients underwent MCGR implantation at the mean age of 7.2 years. The mean follow-up period was 37 months. Fourteen patients (46.7%) underwent an unplanned reoperation within the follow-up period, with a mean time to reoperation of 23 months after initial surgery (range, 5-48 months). Causes of unplanned reoperation were failure of rod distractions, proximal foundation failure, rod breakage, and infection. More frequent distractions (between 1 week and 2 months) were associated with a higher rate of reoperation than distraction frequencies between 3 and 6 months (71% vs. 25%). CONCLUSION: This is the largest series with the longest follow-up to date that examines the need for additional unplanned surgery after the initial procedure. It highlights that MCGR surgery can be associated with unplanned reoperations, and more frequent distractions may be a risk factor. Long-term comparative studies with traditional growing rod are required to evaluate the effectiveness of this implant. LEVEL OF EVIDENCE: 4.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Kwan KYH, Alanay A, Yazici M, et al. Unplanned Reoperations in Magnetically Controlled Growing Rod Surgery for Early Onset Scoliosis With a Minimum of Two-Year Follow-Up. Spine (Phila Pa 1976). 2017. PMID: 28658035. doi:10.1097/BRS.0000000000002297

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。