文献速递关节外科

全髋关节置换术中股骨骨折:第二代锥形楔形股骨柄的设计改进是否降低了发生率?

Total Hip Intraoperative Femur Fracture: Do the Design Enhancements of a Second-Generation Tapered-Wedge Stem Reduce the Incidence?
PubMed 文献 · J Arthroplasty · 2017-05-15

回顾性分析单一术者数据库,比较标准锥形楔形股骨柄与第二代改进型股骨柄在初次非骨水泥全髋关节置换术中术中股骨骨折(IFF)的发生率。标准柄组1510例中41例(2.72%)发生IFF,第二代柄组800例中5例(0.63%)发生IFF。标准柄系统的IFF发生率是第二代柄的4倍以上。

要点

  • 标准锥形楔形股骨柄组IFF发生率为2.72%(41/1510),第二代柄组为0.63%(5/800)。
  • 标准柄系统的IFF发生率比第二代柄高4倍以上。
  • 识别IFF危险因素有助于促进假体系统改进并最大化患者预后。

核心结论

  • 标准锥形楔形股骨柄器械系统在初次非骨水泥全髋关节置换术中导致的IFF发生率是第二代改进型柄的4倍以上。

研究设计与样本

  • 回顾性查询单一经验丰富术者的患者数据库,纳入使用标准锥形楔形股骨柄系统或第二代股骨柄行初次非骨水泥全髋关节置换术的病例。
  • 所有手术均采用保留软组织的解剖囊修复和后路入路。
  • 主要结局指标为IFF,采用比例z检验比较两种股骨柄的IFF差异,并检查患者人口学特征、Dorr分型和假体特征。

主要发现

  • 标准锥形楔形股骨柄组1510例患者中41例(2.72%)发生IFF。
  • 第二代股骨柄组800例患者中5例(0.63%)发生IFF。
  • 未发现其他显著关联。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Intraoperative femur fracture (IFF) is a well-known complication in primary uncemented total hip arthroplasty (THA). Variations in implant instrumentation design and operative technique may influence the risk of IFF. This study investigates IFF between a standard uncemented tapered-wedge femoral stem and its second-generation successor with the following design changes: size-specific medial curvature, proportional incremental stem growth, modest reduction in stem length, and distal lateral relief. METHODS: A single experienced surgeon's patient database was retrospectively queried for IFF occurring during primary uncemented THA using a standard tapered-wedge femoral stem system or a second-generation stem. All procedures were performed using soft tissue preserving anatomic capsule repair and posterior approach. The primary outcome measure was IFF. A z-test of proportions was performed to determine significant difference between the 2 stems with respect to IFF. Patient demographics, Dorr classification, and implant characteristics were also examined. RESULTS: Forty-one of 1510 patients (2.72%) who received a standard tapered-wedge femoral stem sustained an IFF, whereas 5 of 800 patients (0.63%) using the second-generation stem incurred an IFF. No other significant associations were found. CONCLUSION: A standard tapered-wedge femoral stem instrumentation system resulted in greater than 4 times higher incidence of IFF than its second-generation successor used for primary uncemented THA. Identifying risk factors for IFF is necessary to facilitate implant system improvements and thus maximize patient outcomes.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Colacchio ND, Robbins CE, Aghazadeh MS, et al. Total Hip Intraoperative Femur Fracture: Do the Design Enhancements of a Second-Generation Tapered-Wedge Stem Reduce the Incidence? J Arthroplasty. 2017. PMID: 28648706. doi:10.1016/j.arth.2017.05.012

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。