侧方腰椎椎间融合器跨越椎体环状骨突是否能增加失效载荷并减少终板侵犯

Does Spanning a Lateral Lumbar Interbody Cage Across the Vertebral Ring Apophysis Increase Loads Required for Failure and Mitigate Endplate Violation.
PubMed 文献 · Spine (Phila Pa 1976) · 2017-01-01

该随机生物力学尸体研究利用8具新鲜冷冻人脊柱(L1-L5)共40个标本,比较不同尺寸融合器与终板状态对压缩失效和下沉的影响。结果显示跨越环状骨突的长融合器在完整终板上失效载荷增加40%,在去皮质终板上增加30%,且下沉更少。作者认为跨越环状骨突的较大融合器可提高压缩强度并减少下沉,尽管存在终板侵犯或骨质疏松。

要点

  • 跨越环状骨突的长融合器较短的融合器提供更高压缩强度和更少下沉,与终板完整性无关。
  • 完整终板上长融合器(G2)较小融合器(G1)强度显著增加且下沉更少(P=0.003)。
  • 完整终板上长融合器(G2)较去皮质终板上同长度融合器(G4)强度显著增加且抗下沉(P=0.028)。

核心结论

  • 在该骨质疏松尸体模型中,跨越环状骨突使完整终板的失效载荷增加40%,去皮质终板增加30%。
  • 跨越终板环状骨突的较大融合器可提高手术节段压缩强度并减少下沉,尽管存在终板侵犯或骨质疏松。

研究设计与样本

  • 随机生物力学尸体研究,证据等级II。
  • 使用8具新鲜冷冻人脊柱(L1-L5),共40个标本随机分组,在MTS作动器上以5 Hz采集载荷-位移数据直至失效。

主要发现

  • 跨越环状骨突的较长融合器较短的融合器在压缩中提供更大强度且下沉更少,无论终板完整性如何。
  • 完整终板上长融合器(G2)与完整终板上较小融合器(G1)相比,强度显著增加且下沉更少(P=0.003)。
  • 完整终板上长融合器(G2)与去皮质终板上同长度融合器(G4)相比,强度显著增加且抗下沉(P=0.028)。

关键数值与推荐

  • 完整终板上跨越环状骨突的长融合器(G2)与完整终板上较小融合器(G1)比较,强度增加且下沉减少,P=0.003。
  • 完整终板上跨越环状骨突的长融合器(G2)与去皮质终板上同长度融合器(G4)比较,强度增加且抗下沉,P=0.028。
  • 跨越环状骨突使失效载荷在完整终板增加40%,在去皮质终板增加30%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Randomized Biomechanical Cadaveric Study-Level II. OBJECTIVE: We aimed to elucidate that placing lateral lumbar interbody cages that span the stronger ring apophysis will require increasing loads for failure, decreasing rates of subsidence, regardless of bone density or endplate integrity. SUMMARY OF BACKGROUND DATA: There are several reports regarding the rates and grades of cage subsidence when utilizing the lateral lumbar interbody fusion technique. However, there is limited data on how spanning the lateral cage across the ring apophysis can prevent it. METHODS: Eight fresh-frozen human spines (L1-L5) were utilized. Each vertebra was placed with their endplates horizontal in an MTS actuator. A total of 40 specimens were randomized into Groups:Load displacement data was collected at 5 Hz until failure. RESULTS: Longer cages spanning the ring apophysis provided more strength in compression with less subsidence relative to shorter cages, regardless of endplate integrity.Longer cages, spanning the ring apophysis, resting on intact endplates (G2) had a significant (P < 0.05) increase in strength and less subsidence when compared with the smaller cage group resting on intact endplates (G1) (P = 0.003).Longer cages spanning the ring apophysis of intact endplates (G2) showed a significant (P < 0.05) increase in strength and resistance to subsidence when compared with similar length cages resting on decorticated endplates (G4) (P = 0.028). CONCLUSION: Spanning the ring apophysis increased the load to failure by 40% with intact endplates and by 30% with decorticated endplates in this osteoporotic cadaveric model. Larger cages that span the endplate ring apophysis could improve the compressive strength and decrease subsidence at the operative level despite endplate violation or osteoporosis. LEVEL OF EVIDENCE: 2.