文献速递关节外科

相同又不同?全髋关节置换术中的12/14柄与头锥度连接

Same Same but Different? 12/14 Stem and Head Tapers in Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2017-04-27

本研究分析市售12/14柄锥和头锥的设计与表面特征差异。用坐标测量机测几何参数,触觉粗糙度仪测形貌参数。结果显示各厂商锥度在几何和形貌上存在差异,柄锥差异明显,头锥几何不同但形貌无差异,并提出表面光洁度新分类系统。

要点

  • 市售12/14柄锥在几何和形貌参数上因厂商而异。
  • 柄锥的形貌参数变异范围大于几何参数。
  • 头锥几何不同但形貌无差异,提出表面光洁度分类。

核心结论

  • 研究提供了模块化髋锥连接特征与变异的概览,并提出了关于表面光洁度的新分类系统。
  • 这些发现可进一步用于实验腐蚀或取出物研究。

研究设计与样本

  • 对选定的市售12/14柄锥和头锥进行分析。
  • 使用坐标测量机测量锥角、开口锥直径和锥长度等几何参数,使用触觉粗糙度测量仪测定若干形貌参数。

主要发现

  • 所有被研究的锥虽均为所谓12/14锥,但柄锥和头锥因厂商而异。
  • 柄锥在几何和形貌上均明显不同,形貌参数的变异范围大于几何参数。
  • 头锥几何不同,但形貌无差异。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Taper corrosion has been identified to be a major concern in total hip arthroplasty during the past years. So far, the mechanisms that lead to taper corrosion in modular taper junctions are not fully understood. However, it has been shown that corrosion is also influenced by the geometry and topography of the taper, and these parameters vary among the implant manufacturers. The purpose of this study was to investigate the variations of common stem and head tapers regarding design and surface characteristics. METHODS: An analysis of selected commercially available 12/14 stem and head tapers was performed. As geometric parameters, the taper angle, the opening taper diameter, and the taper length were measured using a coordinate measuring machine. Several topographic parameters were determined using a tactile roughness measurement instrument. RESULTS: Although all investigated tapers are so-called 12/14 tapers, this study showed that the stem and head tapers differ among the manufacturers. The stem tapers were clearly different in both geometry and topography, and the range in variation of the topographic parameter was greater than it was for the geometric parameter. In contrast, the head tapers were different in their geometry, although not in topography. CONCLUSION: Ultimately, this study provides an overview on the characteristics and variations of modular hip taper connections, and in addition, a new classification system regarding the surface finish is presented. These findings could be further considered in experimental corrosion or retrieval studies.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Mueller U, Braun S, Schroeder S, et al. Same Same but Different? 12/14 Stem and Head Tapers in Total Hip Arthroplasty. J Arthroplasty. 2017. PMID: 28552447. doi:10.1016/j.arth.2017.04.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。