全髋关节置换术中大型金属股骨头带转接套筒的锥形连接机械稳定性:基于显式有限元模拟的分析
该计算研究采用显式有限元分析,比较50 mm大型金属股骨头配不同长度转接套筒(S、M、L、XL)与32 mm标准股骨头无套筒的锥形连接稳定性,发现带套筒的大直径股骨头微动最高达标准股骨头的7倍,套筒变形随长度增加最高达283%。
要点
- 带转接套筒的大直径股骨头微动最高达无套筒标准股骨头的7倍。
- 套筒变形随长度增加而增大,最高达283%。
- 不同长度套筒的拔出力相当,但接触面积随套筒长度增加而减小。
核心结论
- 该数值研究证实,带转接套筒的大直径股骨头的组装技术相比标准股骨头尺寸表现出更高的微动。
- 股骨柄锥部几何偏差和不正确的手术说明导致锥形连接机械稳定性更差。
研究设计与样本
- 采用显式有限元分析的计算研究,比较50 mm大直径股骨头配不同长度转接套筒(S、M、L、XL)与32 mm标准股骨头无转接套筒的锥形连接稳定性。
- 考虑四种不同冲击配置,由不同锤质量(0.5 kg vs 1.0 kg)和速度(1.0 m/s vs 2.0 m/s)组合产生。
- 通过测定拔出力及模拟行走时关节载荷(分别为2 kN和7.9 Nm)所致的微动来评估锥形稳定性,并评估转接套筒变形和锥形连接接触面积。
主要发现
- 不同尺寸转接套筒的大直径股骨头的拔出力相当。
- 随着转接套筒长度增加,接触面积减小,套筒变形增大(最高达283%)。
- 带转接套筒的大直径股骨头的微动最高达无转接套筒标准股骨头的7倍。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Large diameter heads (LDHs) of metal-on-metal bearings in total hip arthroplasty provide increased range of motion and reduced dislocation rates. However, major concerns grew over high wear rates from the modular connection between femoral stem and head, especially in combination with adapter sleeves. METHODS: A computational study on the taper connection stability of LDH (50 mm) with adapter sleeves of different lengths (S, M, L, and XL) compared with a standard femoral head (32 mm) without adapter sleeves was conducted using explicit finite element analyses. Four different impact configurations were considered resulting from varied mallet mass (0.5 vs 1.0 kg) and velocity (1.0 vs 2.0 m/s). The taper stability was evaluated by determination of the pull-off forces and micromotions due to simulated joint loads during walking (2 kN and 7.9 Nm, respectively). Moreover, the deformations of the adapter sleeves and the contact area in the taper connections were evaluated. RESULTS: Although the pull-off forces of the LDH with different-sized adapter sleeves were comparable, contact area decreased and adapter sleeve deformations increased (up to 283%) with an increasing adapter sleeve length. Moreover, the micromotions of LDH with adapter sleeves were up to 7-times higher, as compared with the standard femoral head without an adapter sleeve. CONCLUSION: The present numerical study confirms that the assembly technique of LDH with adapter sleeves reveals increased micromotions compared with standard femoral head sizes. We could demonstrate that deviations of the stem trunnion geometry and improper surgical instructions led to worse mechanical stability of the taper connection.
引用
- Vogel D, Falkenberg A, Bierbaum S, et al. Mechanical Stability of the Taper Connection of Large Metal Femoral Heads With Adapter Sleeves in Total Hip Arthroplasty Analyzed Using Explicit Finite Element Simulations. J Arthroplasty. 2017. PMID: 28416253. doi:10.1016/j.arth.2017.03.033