翻修全髋关节置换术后两年内假体向近端移位超过1毫米可正确预测80%病例的髋臼假体是否松动:一项术中结局确认的病例对照研究

Proximal translation of > 1 mm within the first two years of revision total hip arthroplasty correctly predicts whether or not an acetabular component is loose in 80% of cases: a case-control study with confirmed intra-operative outcomes.
PubMed 文献 · Bone Joint J · 2017-01-01

本研究回顾性测量84例患者(94个无骨水泥髋臼假体)的EBRA-Cup迁移,比较再翻修术中确认未松动组(52个假体/48例)与松动组(42个假体/36例)。术后3个月起,松动组近端平移和矢状面旋转均显著高于未松动组(p<0.02)。24个月内近端平移>1.0 mm预测松动的阳性预测值为90%、特异度94%、敏感度64%,正确识别76/94(81%)假体的松动状态。

要点

  • 再翻修术前24个月内近端平移>1.0 mm预测髋臼假体松动的阳性预测值为90%、特异度94%、敏感度64%。
  • 术后3个月起,再翻修时确认松动的假体近端平移和矢状面旋转均显著高于未松动假体(p<0.02)。
  • 松动组中10个假体(24%)在术后60个月内近端平移未超过1.0 mm。

核心结论

  • EBRA-Cup测量的近端平移具有高阳性预测值(90%),可用于早期随访识别有无菌性松动风险的患者。
  • 术后60个月内无近端平移提示再翻修全髋关节置换术中假体不太可能松动,但不排除晚期无菌性松动作为失败原因。

研究设计与样本

  • 回顾性病例对照研究,共84例患者(94个无骨水泥髋臼假体),使用EBRA-Cup测量假体迁移。
  • A组为再翻修全髋关节置换术中确认未松动的52个假体(48例),B组为术中确认松动的42个假体(36例)。

主要发现

  • 自术后3个月起,B组平均近端平移和矢状面旋转均显著高于A组(p<0.02)。
  • 24个月内近端平移>1.0 mm预测松动的阳性预测值为90%、特异度94%、敏感度64%。
  • 术后前24个月内近端平移>1.0 mm正确识别94个假体中的76个(81%)为松动或未松动。
  • B组中10个假体(24%)在术后前60个月内近端平移未超过1.0 mm。

关键数值与推荐

  • 近端平移>1.0 mm(24个月内)预测再翻修时髋臼假体松动的阳性预测值为90%、特异度94%、敏感度64%。
  • 近端平移>1.0 mm(24个月内)正确识别76/94(81%)假体的松动状态。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The purpose of this study was to determine the sensitivity, specificity and predictive values of previously reported thresholds of proximal translation and sagittal rotation of cementless acetabular components used for revision total hip arthroplasty (THA) at various times during early follow-up. PATIENTS AND METHODS: Migration of cementless acetabular components was measured retrospectively in 84 patients (94 components) using Ein-Bild-Rontgen-Analyse (EBRA-Cup) in two groups of patients. In Group A, components were recorded as not being loose intra-operatively at re-revision THA (52 components/48 patients) and Group B components were recorded to be loose at re-revision (42 components/36 patients). RESULTS: The mean proximal translation and sagittal rotation were significantly higher in Group B than in Group A from three months onwards (p < 0.02). Proximal translation > 1.0 mm within 24 months had a positive predictive value (PPV) of 90% and a specificity of 94%, but a sensitivity of 64%. Proximal translation > 1.0 mm within the first 24 months correctly identified 76 of 94 (81%) of components to be either loose or not loose. However, ten components in Group B (24%) did not migrate proximally above 1.0 mm within the first 60 months. CONCLUSION: The high PPV of EBRA-Cup measurements of proximal translation (90%) shows that this can be used in early follow-up to identify patients at risk of aseptic loosening. The absence of proximal translation within the first 60 months indicates a component is not likely to be loose at re-revision THA although it does not exclude late aseptic loosening as a cause of failure. Cite this article: Bone Joint J 2017;99-B:465-74.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Kim YS, Abrahams JM, Callary SA, et al. Proximal translation of > 1 mm within the first two years of revision total hip arthroplasty correctly predicts whether or not an acetabular component is loose in 80% of cases: a case-control study with confirmed intra-operative outcomes. Bone Joint J. 2017. PMID: 28385935. doi:10.1302/0301-620X.99B4.BJJ-2016-0805.R1