采用套管准备系统置入部分骨水泥带凹槽肩胛盂假体的早期迁移
研究采用放射立体测量分析(RSA)测量11例骨关节炎患者全肩关节置换术中部分骨水泥带凹槽带栓肩胛盂假体的早期迁移。术后两年随访发现两种迁移模式:一种几乎无迁移且CT显示中心栓周围骨整合,另一种术后三个月内即出现大于6°旋转且CT显示囊性变。11例中3例出现与局灶性透亮区及骨整合缺失相关的快速早期迁移,但两组临床结果相似。
要点
- 11例部分骨水泥带凹槽肩胛盂假体中3例出现快速早期迁移,伴局灶性透亮区及骨整合缺失。
- 早期迁移表现为术后三个月内旋转大于6°,两年CT证实囊性变。
- 无迁移组两年CT显示中心栓周围骨整合,两组临床结果相似。
核心结论
- 11例肩胛盂假体中3例观察到与局灶性透亮区及骨整合缺失相关的快速早期迁移,提示初期稳定性不足导致早期移动和骨整合失败。
研究设计与样本
- 采用放射立体测量分析(RSA)测量11例骨关节炎患者全肩关节置换术中部分骨水泥带凹槽带栓肩胛盂假体的迁移,其中7例男性、4例女性,平均年龄68岁。
- 患者采用改进手术技术置入假体,术后两年进行美国肩肘外科医师(ASES)评分、Constant-Murley评分及CT扫描评估。
主要发现
- 观察到两种迁移模式:第一种几乎无迁移,第二种术后三个月即出现大于6°的旋转。
- 术后两年CT扫描证实两组:一组中心栓周围骨整合,另一组囊性变。
- 中心栓周围骨溶解的患者为早期迁移者,骨整合者早期迁移极小,但两组临床结果相似。
关键数值与推荐
- 早期迁移以术后三个月内旋转大于6°为表现。
- 术后两年CT扫描确认骨整合或囊性变。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Radiostereometric analysis (RSA) allows an extremely accurate measurement of early micromotion of components following arthroplasty. PATIENTS AND METHODS: In this study, RSA was used to measure the migration of 11 partially cemented fluted pegged glenoid components in patients with osteoarthritis who underwent total shoulder arthroplasty using an improved surgical technique (seven men, four women, mean age 68). Patients were evaluated clinically using the American Shoulder and Elbow Surgeons (ASES) and Constant-Murley scores and by CT scans two years post-operatively. RESULTS: There were two patterns of migration, the first showing little, if any, migration and the second showing rotation by > 6° as early as three months post-operatively. At two years, these two groups could be confirmed on CT scans, one with osseointegration around the central peg, and the second with cystic changes. Patients with osteolysis around the central peg were those with early migration and those with osseointegration had minimal early migration. Both groups,however,had similar clinical results. CONCLUSION: Rapid early migration associated with focal lucency and absence of osseointegration was observed in three of 11 glenoid components, suggesting that lack of initial stability leads to early movement and failure of osseointegration. Cite this article: Bone Joint J 2017;99-B:674-9.
引用
- Nuttall D, Birch A, Haines JF, et al. Early migration of a partially cemented fluted glenoid component inserted using a cannulated preparation system. Bone Joint J. 2017. PMID: 28455478. doi:10.1302/0301-620X.99B5.BJJ-2016-0745.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。