影像学标准检测翻修全髋关节置换术后无菌性髋臼假体松动的诊断效能
回顾性分析1980年至2015年间翻修全髋关节置换术后假体,比较EBRA与手动测量在检测无菌性髋臼松动中的诊断效能。EBRA近端平移与矢状旋转的ROC曲线下面积分别为0.94和0.93;2.5 mm近端平移或2°矢状旋转联合放射透亮线标准的敏感性93%、特异性88%。放射透亮线标准单独应用敏感性仅41%、阴性预测值68%。
要点
- EBRA近端平移和矢状旋转是检测无菌性松动的优良诊断试验。
- 2.5 mm近端平移或2°矢状旋转联合放射透亮线标准敏感性93%、特异性88%。
- 放射透亮线标准单独应用敏感性低(41%),不宜单独用于排除松动。
核心结论
- EBRA和手动迁移测量可作为检测翻修全髋关节置换术中非骨水泥髋臼假体无菌性松动的准确诊断工具。
- 鉴于放射透亮线标准敏感性低(41%)和阴性预测值低(68%),不应单独用于排除翻修全髋关节置换术中非骨水泥髋臼假体的松动。
研究设计与样本
- 回顾性研究,纳入1980年至2015年间翻修全髋关节置换术后假体,A组52个假体(48例患者)再翻修时未松动,B组42个假体(36例患者)再翻修时松动。
- 采用EBRA-Cup和手动测量(Sutherland法)评估髋臼假体迁移,并评估放射透亮线的存在与范围。
主要发现
- EBRA近端平移和矢状旋转检测无菌性松动的ROC曲线下面积分别为0.94和0.93。
- EBRA 2.5 mm近端平移或2°矢状旋转联合放射透亮线标准检测无菌性松动的敏感性为93%、特异性为88%。
- 放射透亮线标准的敏感性、特异性、阳性预测值和阴性预测值分别为41%、100%、100%和68%。
- 手动测量近端平移和矢状旋转的ROC曲线下面积分别为0.86和0.92,但手动测量较EBRA特异性降低。
关键数值与推荐
- EBRA近端平移阈值2.5 mm或矢状旋转阈值2°联合放射透亮线标准:敏感性93%、特异性88%。
- 放射透亮线标准:敏感性41%、特异性100%、阳性预测值100%、阴性预测值68%。
- EBRA近端平移ROC曲线下面积0.94,矢状旋转ROC曲线下面积0.93;手动测量近端平移ROC曲线下面积0.86,矢状旋转ROC曲线下面积0.92。
临床怎么用
- EBRA和手动迁移测量可用于检测翻修全髋关节置换术中非骨水泥髋臼假体的无菌性松动。
- 放射透亮线标准不应单独用于排除翻修全髋关节置换术中非骨水泥髋臼假体的松动。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: This study aimed to determine the diagnostic performance of radiographic criteria to detect aseptic acetabular loosening after revision total hip arthroplasty (THA). Secondary aims were to determine the predictive values of different thresholds of migration and to determine the predictive values of radiolucency criteria. PATIENTS AND METHODS: Acetabular component migration to re-revision was measured retrospectively using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements (Sutherland method) in two groups: Group A, 52 components (48 patients) found not loose at re-revision and Group B, 42 components (36 patients) found loose at re-revision between 1980 and 2015. The presence and extent of radiolucent lines was also assessed. RESULTS: Using EBRA, both proximal translation and sagittal rotation were excellent diagnostic tests for detecting aseptic loosening. The area under the receiver operating characteristic (ROC) curves was 0.94 and 0.93, respectively. The thresholds of 2.5 mm proximal translation or 2° sagittal rotation (EBRA) in combination with radiolucency criteria had a sensitivity of 93% and specificity of 88% to detect aseptic loosening. The sensitivity, specificity, positive predictive value and negative predictive value (NPV) of radiolucency criteria were 41%, 100%, 100% and 68% respectively. Manual measurements of both proximal translation and sagittal rotation were very good diagnostic tests. The area under the ROC curve was 0.86 and 0.92 respectively. However, manual measurements had a decreased specificity compared with EBRA. Radiolucency criteria had a poor sensitivity and NPV of 41% and 68% respectively. CONCLUSION: This study shows that EBRA and manual migration measurements can be used as accurate diagnostic tools to detect aseptic loosening of cementless acetabular components used at revision THA. Radiolucency criteria should not be used in isolation to exclude loosening of cementless acetabular components used at revision THA given their poor sensitivity and NPV. Cite this article: Bone Joint J 2017;99-B:458-64.