循环载荷下骨水泥型与非骨水泥型胫骨部件初始稳定性比较
研究将12个骨水泥型与非骨水泥型胫骨底板植入Sawbones模型,模拟体内8周功能的10,000次循环生理性内外侧载荷分布,测量底板周围5个位置的微动。循环载荷对前后、内侧、外侧、后内侧、后外侧托盘边缘最大与最小载荷间微动变化有显著影响;固定技术对前(P < .001)、内(P = .002)、外(P = .0056)侧位置有显著影响,后内侧(P = .36)与后外侧(P = .82)无显著影响。骨水泥型与非骨水泥型部件间微动差异在任何测试位置均未超过150 μm。
要点
- 循环载荷对前后、内侧、外侧、后内侧、后外侧托盘边缘的最大与最小载荷间微动变化有显著影响。
- 固定技术对前(P < .001)、内(P = .002)、外(P = .0056)侧位置微动有显著影响,后内侧(P = .36)与后外侧(P = .82)无显著影响。
- 骨水泥型与非骨水泥型部件间微动差异在任何测试位置均未超过150 μm。
核心结论
- 本研究中非骨水泥型胫骨部件所经历的微动可能提示其初始机械稳定性低于骨水泥型组。
- 然而,在实施的载荷方案下所有测量位置骨水泥型与非骨水泥型平均微动差异均<150 μm,这一初始稳定性差异可能为亚临床水平。
研究设计与样本
- 将12个骨水泥型与非骨水泥型胫骨底板植入Sawbones模型,采用生理性内外侧载荷分布进行10,000次循环测试,以代表体内8周功能。
- 在加载过程中测量底板周围5个位置的微动。
主要发现
- 循环载荷对前后、内侧、外侧、后内侧、后外侧托盘边缘位置的最大与最小载荷间微动变化有显著影响。
- 固定技术对前(P < .001)、内(P = .002)、外(P = .0056)侧位置微动有显著影响,对后内侧(P = .36)与后外侧(P = .82)位置无显著影响。
- 骨水泥型与非骨水泥型部件间微动差异在任何测试位置均未超过150 μm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cement fixation of total knee components remains the gold standard despite resurgence in cementless fixation with the goal of long-term durable fixation. Initial stability is paramount to achieve bony ingrowth of cementless components. METHODS: Twelve cemented and cementless tibial baseplates were implanted into sawbones and tested using a physiological medial-lateral load distribution for 10,000 cycles to represent 8 weeks of in vivo function. Micromotion was measured at 5 locations around the baseplate during loading. RESULTS: Cycling had a significant effect on the change in micromotion between maximum and minimum loads at the anterior, medial, lateral, posteromedial, and posterolateral tray edge locations. A significant effect of fixation technique was detected for the anterior (P < .001), medial (P = .002), and lateral (P = .0056) locations but not for the posteromedial (P = .36) or posterolateral (P = .82) locations. Differences in micromotion between cemented and cementless components did not exceed 150 μm at any tested location. CONCLUSION: The micromotion experienced by cementless tibial components in the present study may indicate a lower initial mechanical stability than the cemented group. However, this difference in initial stability may be subclinical because the differences between average cemented and cementless micromotion were <150 μm at all measured locations under the loading regime implemented.