死后取回膝关节置换假体周围聚乙烯碎屑分布:聚乙烯碎屑能否解释成功全膝关节置换中骨水泥-骨交锁丧失

Peri-Implant Distribution of Polyethylene Debris in Postmortem-Retrieved Knee Arthroplasties: Can Polyethylene Debris Explain Loss of Cement-Bone Interlock in Successful Total Knee Arthroplasties?
PubMed 文献 · J Arthroplasty · 2017-02-03

本研究对8例未翻修死后取回的全膝关节置换(TKA)及2例实验室制备对照,定量骨水泥-骨界面聚乙烯(PE)碎屑分布并与局部交锁丧失比较。PE碎屑可沿界面广泛迁移,但局部骨丢失与PE碎屑量不相关,提示其他因素可能解释交锁丧失。

要点

  • 死后取回TKA的PE颗粒数密度范围为8.6(1.3)至24.9(3.1)颗粒/mm²(标准误)。
  • 界面远端(>5mm)平均颗粒数密度是界面处的两倍。
  • 界面局部交锁丧失与PE存在与否或颗粒密度无关。

核心结论

  • PE碎屑可沿固定良好的胫骨假体骨水泥-骨界面广泛迁移。
  • 界面局部骨丢失与PE碎屑量不相关,提示观察到的交锁丧失可能源于其他因素。

研究设计与样本

  • 研究纳入8例新鲜、未翻修的死后取回全膝关节置换(TKA),并以2例实验室制备构建体作为阴性对照。
  • 将每例胫骨近端完整骨水泥-骨界面包埋于Spurr树脂中,切片后在偏振光下成像以识别双折射PE颗粒。
  • 定量骨水泥-骨界面及界面远端的PE磨损颗粒数密度,并与局部骨水泥-骨交锁丧失比较。

主要发现

  • 死后取回TKA的平均PE颗粒数密度范围为8.6(1.3)至24.9(3.1)颗粒/mm²(标准误),但与使用年限仅弱相关。
  • 界面远端(>5mm)的平均颗粒数密度是界面处的两倍。
  • 界面局部交锁丧失与PE的存在、缺失或颗粒密度无关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Loss of mechanical interlock between cement and bone with in vivo service has been recently quantified for functioning, nonrevised, cemented total knee arthroplasties (TKAs). The cause of interlocking trabecular resorption is not known. The goal of this study is to quantify the distribution of PE debris at the cement-bone interface and determine if polyethylene (PE) debris is locally associated with loss of interlock. METHODS: Fresh, nonrevised, postmortem-retrieved TKAs (n = 8) were obtained en bloc. Laboratory-prepared constructs (n = 2) served as negative controls. The intact cement-bone interface of each proximal tibia was embedded in Spurr's resin, sectioned, and imaged under polarized light to identify birefringent PE particles. PE wear particle number density was quantified at the cement-bone interface and distal to the interface, and then compared with local loss of cement-bone interlock. RESULTS: The average PE particle number density for postmortem-retrieved TKAs ranged from 8.6 (1.3) to 24.9 (3.1) particles/mm2 (standard error) but was weakly correlated with years in service. The average particle number density was twice as high as distal (>5mm) to the interface compared to at the interface. The local loss of interlock at the interface was not related to the presence, absence, or particle density of PE. CONCLUSION: PE debris can migrate extensively along the cement-bone interface of well-fixed tibial components. However, the amount of local bone loss at the cement-bone interface was not correlated with the amount of PE debris at the interface, suggesting that the observed loss of trabecular interlock in these well-fixed TKAs may be due to alternative factors.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Cyndari KI, Goodheart JR, Miller MA, et al. Peri-Implant Distribution of Polyethylene Debris in Postmortem-Retrieved Knee Arthroplasties: Can Polyethylene Debris Explain Loss of Cement-Bone Interlock in Successful Total Knee Arthroplasties? J Arthroplasty. 2017. PMID: 28285038. doi:10.1016/j.arth.2017.01.047