文献速递关节外科

体内三维髌骨力学:正常膝关节与穹顶形及解剖形髌骨假体的比较

In Vivo Three-Dimensional Patellar Mechanics: Normal Knees Compared with Domed and Anatomic Patellar Components.
PubMed 文献 · J Bone Joint Surg Am · 2017-01-01

本研究在负重深屈膝过程中,用荧光透视结合CT与CAD模型三维配准,比较10例正常膝、10例LCS-PS解剖形髌骨假体膝和10例PFC Sigma RP-PS穹顶形髌骨假体膝的髌股运动学与接触力学。两种假体设计均表现出良好的旋转运动学并在股骨滑车内良好追踪,但接触面积均小于既往报道的正常髌骨。

要点

  • 两种髌骨假体设计均表现出良好旋转运动学,并在股骨滑车内良好追踪。
  • LCS-PS组接触面积在屈膝60°达峰值201±63.4 mm,Sigma RP-PS组为218±95.4 mm。
  • TKA组接触点保持在髌骨假体中心附近。

核心结论

  • 两种髌骨假体设计均表现满意,但与正常髌骨相比髌股接触面积减小。

研究设计与样本

  • 研究纳入10例正常膝、10例接受LCS-PS全膝关节置换(解剖形髌骨假体)的膝和10例接受PFC Sigma RP-PS全膝关节置换(穹顶形髌骨假体)的膝。
  • 在荧光透视监测下,患者完成从完全伸膝到最大屈膝的负重深屈膝动作。
  • 从CT扫描分割骨几何形态,从制造商获取植入假体的计算机辅助设计(CAD)模型。
  • 采用三维到二维配准过程获取三维髌股运动学。
  • 使用髌骨与股骨关节面之间的距离图计算接触力学。

主要发现

  • 与正常髌骨相比,两种髌骨假体设计均表现出良好旋转运动学,并在股骨滑车内良好追踪。
  • TKA组接触面积在屈膝60°达峰值,LCS-PS组为201±63.4 mm,Sigma RP-PS组为218±95.4 mm。
  • TKA组接触面积明显小于既往报道的正常髌骨接触面积。
  • TKA组接触点保持在髌骨假体中心附近。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Patellofemoral complications are a major cause of revision surgery following total knee arthroplasty (TKA). High forces occurring at the patellofemoral articulation coupled with a small patellofemoral contact area pose substantial design challenges. In this study, the three-dimensional (3D) in vivo mechanics of domed and anatomically shaped patellar components were compared with those of native patellae. METHODS: Ten normal knees, 10 treated with an LCS-PS (low contact stress-posterior stabilized) TKA (anatomically shaped patellar component), and 10 treated with a PFC Sigma RP-PS (press-fit condylar Sigma rotating platform-posterior stabilized) TKA (domed patellar component) were analyzed under fluoroscopic surveillance while the patient performed a weight-bearing deep knee bend from full knee extension to maximum knee flexion. Relevant bone geometries were segmented out from computed tomography (CT) scans, and computer-assisted-design (CAD) models of the implanted components were obtained from the manufacturer. Three-dimensional patellofemoral kinematics were obtained using a 3D-to-2D registration process. Contact mechanics were calculated using a distance map between the articulating patellar and femoral surfaces. RESULTS: Both patellar component designs exhibited good rotational kinematics and tracked well within the femoral trochlea when compared with the normal patella. The contact areas in the TKA groups peaked at 60° of knee flexion (mean and standard deviation, 201 ± 63.4 mm for the LCS-PS group and 218 ± 95.4 mm for the Sigma RP-PS group), and the areas were substantially smaller than those previously reported for the normal patella. Contact points in the TKA groups stayed close to the center of the patellar components. CONCLUSIONS: Both designs performed satisfactorily, although patellofemoral contact areas were reduced in comparison with those in the native patella. LEVEL OF EVIDENCE: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Sharma A, Grieco TF, Zingde SM, et al. In Vivo Three-Dimensional Patellar Mechanics: Normal Knees Compared with Domed and Anatomic Patellar Components. J Bone Joint Surg Am. 2017. PMID: 28244917. doi:10.2106/JBJS.15.01095

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。