模块化钴铬髋臼内衬背侧的微动与腐蚀:回收分析

Fretting and Corrosion at the Backside of Modular Cobalt Chromium Acetabular Inserts: A Retrieval Analysis.
PubMed 文献 · J Arthroplasty · 2016-10-04

对18个回收的模块化双动(MDM)钴铬髋臼内衬与30个金属对金属(MoM)全髋关节置换的同类模块化内衬进行背侧微动和腐蚀评估。MDM内衬未发现特定微动或腐蚀模式,MoM组微动和腐蚀显著更重,尤其在锥度啮合区。MDM内衬缺乏显著微动和腐蚀,与近期认为该界面可产生临床显著金属沉着和不良局部组织反应的说法不一致。

要点

  • 18个MDM内衬背侧未见特定微动或腐蚀模式。
  • MoM组微动和腐蚀评分显著高于MDM组,差异来自啮合锥度区。
  • MDM内衬的发现与近期金属沉着和不良局部组织反应的说法不一致。

核心结论

  • MoM组件比MDM设计表现出更多微动和腐蚀,特别是在锥度区,可能由内衬设计的锥度啮合机制和几何形状差异驱动。
  • MDM内衬中未观察到显著微动和腐蚀,与近期认为该界面可产生临床显著金属沉着和不良局部组织反应的说法不一致。

研究设计与样本

  • 回收分析,评估18个回收的钴铬MDM髋臼内衬背侧,并以30个金属对金属全髋关节置换的同类模块化内衬作为对照。
  • 基于先前建立的方法评估极区和锥度区的微动和腐蚀。

主要发现

  • MDM内衬上未发现特定的微动或腐蚀模式。
  • MoM队列的微动和腐蚀均显著大于MDM队列,由MoM内衬啮合锥度区更高的微动和腐蚀评分驱动。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Adverse local tissue reaction formation has been suggested to occur with the Modular Dual Mobility (MDM) acetabular design. Few reports in the literature have evaluated fretting and corrosion damage between the acetabular shell and modular metal inserts in this modular system. We evaluated a series of 18 retrieved cobalt chromium MDM inserts for evidence of fretting and corrosion. METHODS: We assessed the backsides of 18 MDM components for evidence of fretting and corrosion in polar and taper regions based on previously established methods. We collected and assessed 30 similarly designed modular inserts retrieved from metal-on-metal (MoM) total hip arthroplasties as a control. RESULTS: No specific pattern of fretting or corrosion was identified on the MDM inserts. Both fretting and corrosion were significantly greater in the MoM cohort than the MDM cohort, driven by higher fretting and corrosion scores in the engaged taper region of the MoM inserts. CONCLUSION: MoM components demonstrated more fretting and corrosion than MDM designs, specifically at the taper region, likely driven by differences in the taper engagement mechanism and geometry among the insert designs. The lack of significant fretting and corrosion observed in the MDM inserts are inconsistent with recent claims that this interface may produce clinically significant metallosis and adverse local tissue reactions.

来源:J Arthroplasty,查看原文。版权归原作者所有。