Morscher 压配式髋臼假体:18-22 年独立长期回顾
本研究回顾 122 例(111 例患者)使用 Morscher 一体式压配髋臼假体的全髋关节置换,平均随访 19.7 年。20 年 Kaplan-Meier 生存率全因为翻修 93.4%,任何髋臼翻修 95.5%,髋臼无菌性松动、磨损或骨溶解为 97.1%。作者认为该假体在 20 年时仍表现良好。
要点
- 平均随访 19.7 年,全因翻修率每 100 个观察假体年 0.32 次。
- 20 年全因翻修生存率 93.4%(95%CI 86.6-96.8)。
- 20 年髋臼无菌性松动、磨损或骨溶解生存率 97.1%(95%CI 91.2-99.1)。
核心结论
- Morscher 髋臼假体在使用传统聚乙烯的情况下 20 年时仍表现良好,结果与其他非骨水泥髋臼假体相当或更优。
研究设计与样本
- 回顾性长期随访研究,共 122 例全髋关节置换(111 例患者),使用 Morscher 一体式压配髋臼假体,平均随访 19.7 年。
主要发现
- 22 例患者(25 髋)死亡,7 髋接受翻修,其中 5 髋为髋臼翻修;6 例患者(6 髋)拒绝参与但已知未翻修。
- 平均 Oxford 髋关节评分 41.1(范围 22-48),平均简化 Western Ontario 与 McMaster 大学骨关节炎指数评分 5.7/48(范围 0-24)。
- 82 个有影像学资料的存活髋中,偏心性磨损 13 个(15.7%),严重骨溶解 14 个(17%)。
- 全因翻修率为每 100 个观察假体年 0.32 次(95%CI 0.13-0.66)。
- 20 年 Kaplan-Meier 生存率:全因翻修 93.4%(95%CI 86.6-96.8),任何髋臼翻修 95.5%(95%CI 89.4-98.1),髋臼无菌性松动、磨损或骨溶解 97.1%(95%CI 91.2-99.1)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There are relatively few 20-year results of uncemented acetabular components, and most of these are modular designs. This study reports the 20-year results of a monoblock press-fit acetabular component. METHODS: A total of 122 total hip arthroplasties (111 patients) using the Morscher cup were reviewed at a mean of 19.7 years. The average age at implantation was 57.3 years (range, 36-74 years), and 81 (66%) were men. RESULTS: Twenty-two patients (25 hips) had died. Seven hips were revised, including 5 acetabular revisions. Six patients (6 hips) declined to participate but were known not to have been revised. The mean Oxford hip score was 41.1 (range, 22-48), and the mean reduced Western Ontario and McMaster Universities Osteoarthritis Index score was 5.7/48 (range, 0-24). Eccentric wear was seen in 13 (15.7%) and major osteolysis in 14 (17%) of 82 surviving hips with radiographs. The all-cause revision rate was 0.32 per 100 observed component years (95% confidence interval [CI], 0.13-0.66). The 20-year Kaplan-Meier survival was 93.4% (CI, 86.6-96.8) for all-cause revisions, 95.5% (CI, 89.4-98.1) for any acetabular revision, and 97.1% (CI, 91.2-99.1) for acetabular aseptic loosening, wear, or osteolysis. CONCLUSION: The Morscher acetabular component has continued to perform well at 20 years despite using conventional polyethylene with results that match or surpass other cementless acetabulae.
引用
- Gwynne-Jones DP, Lash HWR, James AW, et al. The Morscher Press-Fit Acetabular Component: An Independent Long-Term Review at 18-22 Years. J Arthroplasty. 2017. PMID: 28343828. doi:10.1016/j.arth.2017.02.052
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。