模块化髋关节假体颈干耦合部件的取出分析

Retrieval Analysis of Neck-Stem Coupling in Modular Hip Prostheses.
PubMed 文献 · J Arthroplasty · 2017-02-14

本研究对60例取出的Rejuvenate模块化颈干假体与26例来自7种其他模块化颈设计假体的磨损和腐蚀行为进行比较,并回顾组织学ALVAL分级。结果显示Rejuvenate设计的损伤和腐蚀评分更高(P<.01),组织学ALVAL可能性为其他设计的20倍(P<.01);混合金属耦合的微动磨损(P<.01)和腐蚀(P=.02)评分高于非混合金属耦合。作者建议临床医生避免植入模块化颈干系统。

要点

  • Rejuvenate设计的损伤和腐蚀评分高于其他7种设计(P<.01),组织学ALVAL可能性为其他设计的20倍(P<.01)。
  • 混合金属耦合的微动磨损评分(P<.01)和腐蚀评分(P=.02)均高于非混合金属耦合。
  • 所有模块化颈干取出假体均发生微动磨损和腐蚀,与设计无关;作者建议临床医生避免植入模块化颈干系统。

核心结论

  • 所有模块化颈干取出假体均发生微动磨损和腐蚀,与设计无关;但混合金属耦合比同质金属耦合遭受更多腐蚀。
  • 作者认为这可能源于钛合金柄较低的模量,使其在与钴合金模块化颈对合加载时金属转移和表面损伤增加,进而可能解释更高的ALVAL和腐蚀评分。
  • 鉴于混合金属增加腐蚀风险、非混合金属柄颈增加颈骨折风险,作者建议临床医生避免植入模块化颈干系统。

研究设计与样本

  • 本研究比较了60例取出的Rejuvenate模块化颈干假体与26例来自7种其他模块化颈设计假体的微动磨损和腐蚀行为。
  • 对26例额外病例的翻修时组织学切片进行审查,并按无菌性淋巴细胞主导性血管炎相关病变(ALVAL)分级。
  • 进行多变量分析以评估微动磨损和腐蚀差异,并校正混杂因素(如植入时长)。

主要发现

  • Rejuvenate设计的损伤和腐蚀评分高于其他7种设计(P<.01)。
  • 召回设计的组织学样本显示ALVAL的可能性是其他设计样本的20倍(P<.01)。
  • 混合金属耦合的微动磨损评分高于非混合金属耦合(P<.01)。
  • 混合金属耦合的腐蚀评分高于非混合金属耦合(P=.02)。

临床怎么用

  • 作者建议临床医生避免植入模块化颈干系统。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Dual-taper modular stems have suffered from high revision rates caused by adverse local tissue reactions secondary to fretting and corrosion. We compared the fretting and corrosion behavior of a group of modular neck designs to that of a design that had been recalled for risks associated with fretting and corrosion at the modular neck junction. METHODS: We previously analyzed fretting and corrosion on 60 retrieved Rejuvenate modular neck-stem implants. Here we compare those results to results from 26 retrieved implants from 7 other modular neck designs. For the 26 additional cases, histology slides of tissue collected at revision were reviewed and graded for aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Multivariate analyses were performed to assess differences in fretting and corrosion, adjusting for confounding factors (eg, length of implantation). RESULTS: The Rejuvenate design had higher damage and corrosion scores than the other 7 designs (P < .01). Histologic samples from the recalled design were 20 times more likely to show ALVAL than samples from the other designs (P < .01). Mixed metal couples had higher fretting (P < .01) and corrosion (P = .02) scores than non-mixed metal couples. CONCLUSION: Fretting and corrosion occurred on all modular neck-stem retrievals regardless of design. However, mixed metal couples suffered more corrosion than homogenous couples. This may be due to the lower modulus of the titanium alloy used for the stem, allowing for increased metal transfer and surface damage when loaded against a cobalt alloy modular neck, which in turn could account for the higher ALVAL and corrosion scores. Due to increased corrosion risk with mixed metals and increased neck fracture risk with non-mixed metal stem and necks, we suggest that clinicians avoid implantation of modular neck-stem systems.