使用撑开器是韧带平衡的准确方法吗?

Is the Use of Spreaders an Accurate Method for Ligament Balancing?
PubMed 文献 · J Arthroplasty · 2017-02-07

研究在22具新鲜冷冻尸体膝关节上由22名外科医生行全膝关节置换术,比较徒手撑开器目视法与盲法及控制张力器测量伸直和屈曲间隙的准确性。结果显示撑开器法测得的伸直和屈曲间隙均偏大且不对称(P < .0001),目视法与盲法之间无视觉影响。

要点

  • 撑开器徒手施力测得的伸直和屈曲间隙较张力器法偏大且不对称(P < .0001)。
  • 目视法与盲法在撑开器测量中未见视觉影响。
  • 目视法间隙偏大和不对称程度略小于盲法,但差异显著(P < .0001)。

核心结论

  • 与使用张力器相比,经验性徒手施力撑开器评估伸直和屈曲间隙产生偏大且不对称的间隙。
  • 与盲法评估相比,撑开器施加经验性徒手力期间未观察到视觉影响。

研究设计与样本

  • 使用22具新鲜冷冻尸体膝关节,由22名外科医生行全膝关节置换术。
  • 伸直和屈曲间隙以撑开器经验性徒手施力按两种方式测量:外科医生对间隙几何形状盲法(盲法组BM)和外科医生目视(目视组VM)。
  • 使用张力器测量撑开器测量期间施加的相应韧带张力,并以标准力100 N和80 N测量每个股胫间室的伸直和屈曲间隙(张力器法TM)。

主要发现

  • 所有撑开器测量(VM和BM)与张力器测量的相同间隙相比,伸直和屈曲间隙均偏大且不对称(P < .0001)。
  • VM组约63%(P < .001)和77.3%(P = .161),BM组68.2%(P = .018)和77.3%(P = .161)的伸直和屈曲间隙与TM相比不对称达3 mm。
  • 与TM相比,VM组测量的间隙偏大和不对称程度略小于BM组,但差异显著(P < .0001)。

关键数值与推荐

  • 张力器法以标准力100 N和80 N测量每个股胫间室的伸直和屈曲间隙。
  • VM组约63%(P < .001)和77.3%(P = .161)的伸直和屈曲间隙与TM相比不对称达3 mm。
  • BM组68.2%(P = .018)和77.3%(P = .161)的伸直和屈曲间隙与TM相比不对称达3 mm。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To analyze 2 methods of manual spreader gap assessment accuracy, visual vs blinded, compared with a controlled tensioner in total knee arthroplasty. METHODS: Twenty-two fresh frozen cadaver knees were used to perform total knee arthroplasty by 22 surgeons. Extension and flexion gaps were measured with empirical manual force application with spreaders in 2 different manners: (1) surgeons were blinded to gap geometry formation-blind method group (BM) and (2) surgeons viewed them-viewing method group (VM). A tensioner was used to measure the corresponding ligament tension applied during spreader measurements and to measure the extension and flexion gaps with standard force of 100 and 80 N (tensioner method [TM]) in each femorotibial compartment. RESULTS: All measurements with spreaders (VM and BM) presented extension and flexion gaps oversized and asymmetric (P < .0001), when compared with the same gaps measured with the tensioner. Approximately 63% (P = <0.001) and 77.3% (P = .161) of the VM group and 68.2% (P = .018) and 77.3% (P = .161) of the BM group demonstrated asymmetry for extension and flexion gaps up to 3 mm to the TM. Gaps measured in the VM group presented results with slightly less oversizing and asymmetries than the measurements in the BM group compared with TM, although significantly different (P < .0001). CONCLUSION: The assessment of extension and flexion gaps with empirical manual applied force spreaders produced oversized and asymmetric gaps compared with the use of tensioner. No visual influence was observed during the spreader applied empirical manual force compared with the blinded assessment.