仪表化膝关节受力能否代表更广泛的后交叉韧带保留型全膝关节置换术人群?
本研究纳入23例后交叉韧带保留型全膝关节置换术患者,通过运动分析结合计算模型计算上下楼梯和坐椅站起活动的轴向假体力。结果显示楼梯上升和下降时人群峰值力与仪表化假体报告值存在差异(P=.03),但变异性无差异(P=.106至P=.99),作者认为该数据可用于位移驱动磨损测试。
要点
- 楼梯上升峰值轴向力为3.06 kN,楼梯下降为2.74 kN,坐椅站起为2.65 kN,坐椅坐下为2.60 kN。
- 人群与仪表化假体峰值力在活动周期后半段楼梯上升(P=.03)和下降(P=.03)存在差异。
- 峰值力和冲量的变异性无差异(P=.106至P=.99)。
核心结论
- 本研究提出的力和运动数据代表后交叉韧带保留型全膝关节置换术患者,可用于位移驱动的膝关节磨损测试。
- 本研究测得的力与文献中后交叉韧带牺牲型超 congruency 假体的仪表化假体所测力相似。
研究设计与样本
- 本研究为机构审查委员会批准的、经知情同意后开展的23例后交叉韧带保留型全膝关节置换术受试者的运动分析研究。
- 采用先前验证过的计算模型计算轴向全膝关节置换术力,并用t检验和Levene检验比较人群与仪表化假体峰值力和冲量的均值及变异性。
主要发现
- 楼梯上升、楼梯下降、坐椅站起和坐椅坐下的峰值轴向力分别为3.06、2.74、2.65和2.60 kN。
- 楼梯上升、楼梯下降、坐椅坐下和坐椅站起的力冲量分别为123.3、123.4、153.5和154.0 kN*%活动周期。
- 在活动周期后半段,人群全膝关节置换术与仪表化全膝关节置换术的峰值力在楼梯上升(P=.03)和楼梯下降(P=.03)存在差异。
- 峰值力和冲量的变异性无差异(P=.106至P=.99)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: It is not known if the loads and motions reported for instrumented knees are generalizable to a larger population of total knee arthroplasty (TKA) patients. The purpose of this study is to (1) report axial implant force data for chair and stair activities for a population of cruciate-retaining TKA patients and (2) compare the population forces to those measured with instrumented TKAs. METHODS: Twenty-three subjects with a cruciate-retaining TKA underwent motion analysis during stair ascending, stair descending, chair sitting, and chair rising activities after informed consent in this institutional review board approved study. Axial TKA forces were calculated using a previously validated computational model. Differences between the mean and variability of population instrumented TKA peak forces and force impulses were tested using t tests and Levene test. RESULTS: Peak axial forces were 3.06, 2.74, 2.65, and 2.60 kN for stair ascent, stair descent, chair rising, and chair sitting, respectively. Force impulses were 123.3, 123.4, 153.5, and 154.0 kN*% activity cycle for stair ascent, stair descent, chair sitting, and chair rising, respectively. Population TKA and instrumented TKA peak forces were different for stair ascent (P = .03) and stair descent (P = .03) in the second half of the activity cycles. The variability of the peak forces and impulses were not different (P = .106 to P = .99). CONCLUSION: The forces and motions presented in this study represent cruciate-retaining TKA patients and could be used for displacement-driven knee wear testing. The forces are similar to those in the literature from instrumented prostheses of an ultracongruent cruciate-sacrificing TKA.
引用
- Freed RD, Simon JC, Knowlton CB, et al. Are Instrumented Knee Forces Representative of a Larger Population of Cruciate-Retaining Total Knee Arthroplasties? J Arthroplasty. 2017. PMID: 28262455. doi:10.1016/j.arth.2017.01.054
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。