回收的组配式全膝关节置换假体的腐蚀、微动磨损与材料损失评估
本研究对同一制造商的22例组配式全膝关节翻修取出的假体进行定性视觉评分和定量材料损失测量,评估锥度连接界面(轴向螺钉或径向螺钉)在钴铬合金与钛合金组合下的腐蚀与微动磨损。结果显示混合金属界面中钴铬组件最大线性深度高于钛组件,钛合金组件微动磨损评分更高;总体材料损失低,无因临床金属相关反应而翻修。
要点
- 混合金属界面中钴铬组件最大线性深度高于对应钛组件(P = .007)。
- 混合金属界面中钛合金组件微动磨损评分统计学高于其余组。
- 总体材料损失低,无因临床金属相关反应而翻修。
核心结论
- 结果提示混合金属界面中的钴铬组件比其他组件更易发生腐蚀,与钛合金接触时可能存在优先腐蚀。
- 尽管本系列中观察到腐蚀,但材料损失低,且无因临床金属相关反应而翻修,提示全膝关节置换中腐蚀的临床影响低。
研究设计与样本
- 从经机构审查委员会批准的数据库中识别同一制造商的22例组配式全膝关节翻修取出假体。
- 连接界面设计包括带轴向螺钉的锥度和带径向螺钉的锥度,构造由钴铬合金和钛合金两种金属合金组成。
- 对组件进行腐蚀和微动磨损的定性评分与定量测量,负值代表附着材料;使用符号检验分析统计差异,使用Spearman秩相关检验分析相关性(P < .05)。
主要发现
- 总体人群的中位体积材料损失为-0.23 mm³,最大线性深度为5.84 μm。
- 混合金属界面中钴铬组件的最大线性深度高于对应钛组件(P = .007)。
- 混合金属界面中钛合金组件的微动磨损评分统计学高于其余组。
- 锥度角度与材料损失无相关性。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Modular junctions in total hip arthroplasties have been associated with fretting, corrosion, and debris release. The purpose of this study is to analyze damage severity in total knee arthroplasties of a single design by qualitative visual assessment and quantitative material loss measurements to evaluate implant performance and patient impact via material loss. METHODS: Twenty-two modular knee retrievals of the same manufacturer were identified from an institutional review board-approved database. Junction designs included tapers with an axial screw and tapers with a radial screw. Constructs consisted of 2 metal alloys: CoCr and Ti6Al4V. Components were qualitatively scored and quantitatively measured for corrosion and fretting. Negative values represent adhered material. Statistical differences were analyzed using sign tests. Correlations were tested with a Spearman rank order test (P < .05). RESULTS: The median volumetric material loss and the maximum linear depth for the total population were -0.23 mm3 and 5.84 μm, respectively. CoCr components in mixed metal junctions had higher maximum linear depth (P = .007) than corresponding Ti components. Fretting scores of Ti6Al4V alloy components in mixed metal junctions were statistically higher than the remaining groups. Taper angle did not correlate with material loss. CONCLUSION: Results suggest that CoCr components in mixed metal junctions are more vulnerable to corrosion than other components, suggesting preferential corrosion when interfacing with Ti6Al4V. Overall, although corrosion was noted in this series, material loss was low, and none were revised for clinical metal-related reaction. This suggests the clinical impact from corrosion in total knee arthroplasty is low.
引用
- Martin AJ, Seagers KA, Van Citters DW, et al. Assessment of Corrosion, Fretting, and Material Loss of Retrieved Modular Total Knee Arthroplasties. J Arthroplasty. 2017. PMID: 28343824. doi:10.1016/j.arth.2017.02.047