通过检查未使用和取出的假体研究当代金属对金属髋关节置换系统中的锥度失效

Investigation of Taper Failure in a Contemporary Metal-on-Metal Hip Arthroplasty System Through Examination of Unused and Explanted Prostheses.
PubMed 文献 · J Bone Joint Surg Am · 2017-01-01

本研究分析93个与Corail钛合金非骨水泥股骨柄配合使用的36mm轴承直径Articul/eze股骨头锥度,采用多元回归模型评估母锥面制造表面形貌对锥度磨损的影响。发现母锥面制造表面粗糙度(Ra值范围0.14至4.20μm)是与锥度磨损相关的最重要变量(p<0.001),最佳拟合模型解释约44%的锥度磨损率变异。结论认为母锥面粗糙度似乎是头-柄锥度连接处金属碎屑释放的重要因素。

要点

  • 母锥面制造表面粗糙度是与锥度磨损相关的最重要变量(p<0.001)。
  • 最佳拟合回归模型(含体内时间、头偏移、Rpk值、柄轴角度和轴承面磨损率)解释约44%的锥度磨损率变异。
  • 母锥面制造表面光洁度存在很大变异,Ra值范围为0.14至4.20μm。

核心结论

  • 母锥面粗糙度似乎是头-柄锥度连接处金属碎屑释放的重要因素。
  • 制造工艺中先前未被认识的变异可能对患者临床结局产生重大影响。

研究设计与样本

  • 研究分析了93个与Corail钛合金非骨水泥股骨柄配合使用的36mm轴承直径Articul/eze股骨头锥度。
  • 采用多元回归模型评估母锥面制造表面形貌对锥度磨损的影响,并考虑其他已知变量。

主要发现

  • 母锥面制造表面光洁度存在很大变异,测量Ra值范围为0.14至4.20μm。
  • 母锥面粗糙度是与锥度磨损相关的最重要变量(p<0.001)。
  • 最佳拟合回归模型包括体内时间、头偏移、Rpk值、柄轴角度和轴承面磨损率,解释约44%的锥度磨损率变异。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Large-diameter (≥36-mm) metal-on-metal (MoM) total hip replacements have been shown to fail at an unacceptably high rate. Globally, the DePuy Pinnacle prosthesis was the most widely used device of this type. There is evidence to suggest that one of the main reasons for the poor clinical performance of large-diameter MoM prostheses is the metal debris released from the head-stem taper junction-i.e., taper junction failure. The aim of this study was to investigate variation in the as-manufactured finish of the female taper surface and to determine its influence on material loss. We hypothesized that rougher surfaces with higher relative material peaks would be significantly associated with greater taper wear rates. METHODS: We analyzed 93 Articul/eze femoral head tapers with a 36-mm bearing diameter that had been used in combination with a Corail titanium uncemented stem. The influence of the surface topography of the as-manufactured female taper surface on taper wear was examined by means of a multiple regression model, taking into account other known variables. RESULTS: We identified great variation in the as-manufactured surface finish of the female taper surface, with a range of measured Ra values from 0.14 to 4.20 μm. The roughness of the female taper surface appeared to be the most important variable associated with taper wear (p < 0.001). The best-fitting regression model, including duration in vivo, head offset, reduced peak height (Rpk) value, stem shaft angle, and bearing surface wear rate, explained approximately 44% of the variation in taper wear rates. CONCLUSIONS: We concluded that the roughness of the female taper surface appears to be a significant factor in metal debris release from head-stem taper junctions. CLINICAL RELEVANCE: This study shows evidence that previously unappreciated variations in manufacturing processes may have a major impact on the clinical outcomes of patients.