新型多孔钛干骺端锥翻修全膝关节置换术的研制与验证

Development and Verification of Novel Porous Titanium Metaphyseal Cones for Revision Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2017-01-18

本研究基于CT解剖数据库设计并3D打印新型多孔钛干骺端锥,在台架试验模型中比较其与常用多孔钽锥在生理载荷下的微动。新型钛股骨锥在12个方向中的5个方向微动显著小于传统钽锥(P < .05),胫骨锥在内翻/外翻(P = .004)和后部压缩微动(P = .002)方面小于传统系统。结论认为该解剖型多孔钛锥具有满意的机械稳定性,但需长期临床随访。

要点

  • 新型多孔钛股骨锥在12个测量方向中的5个方向微动显著小于传统钽锥(P < .05),其余7个方向无统计学差异。
  • 新型多孔钛胫骨锥在内翻/外翻(P = .004)和后部压缩微动(P = .002)方面小于传统多孔钽系统。
  • 新型钛锥稳定性等效或优于钽锥,但需长期临床随访。

核心结论

  • 生物力学研究表明,基于解剖设计的多孔钛干骺端锥系统在生理载荷下测量微动,对股骨和胫骨骨缺损具有满意的机械稳定性。
  • 新型锥设计与提高手术效率的器械相结合令人鼓舞。
  • 需进行长期临床随访。

研究设计与样本

  • 基于CT解剖数据库的股骨和胫骨形态设计新型多孔钛干骺端锥,并采用3D打印技术制造。
  • 在台架试验模型中,比较新型钛锥与广泛使用的多孔钽干骺端锥系统在生理载荷条件下的稳定性。

主要发现

  • 新型钛股骨锥构建体在12个测量方向中的5个方向微动显著小于传统股骨锥构建体(P < .05),7个方向无统计学差异。
  • 新型多孔钛胫骨干骺端锥在内翻/外翻(P = .004)和后部压缩微动(P = .002)方面小于传统多孔钽系统。
  • 新型钛锥的稳定性等效或优于钽锥。

局限性

  • 需进行长期临床随访。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Porous metaphyseal cones are widely used in revision knee arthroplasty. A new system of porous titanium metaphyseal cones has been designed based on the femoral and tibial morphology derived from a computed tomography-based anatomical database. The purpose of this study is to evaluate the initial mechanical stability of the new porous titanium revision cone system by measuring the micromotion under physiologic loading compared with a widely-used existing porous tantalum metaphyseal cone system. METHODS: The new cones were designed to precisely fit the femoral and tibial anatomy, and 3D printing technology was used to manufacture these porous titanium cones. The stability of the new titanium cones and the widely-used tantalum cones were compared under physiologic loading conditions in bench top test model. RESULTS: The stability of the new titanium cones was either equivalent or better than the tantalum cones. The new titanium femoral cone construct had significantly less micromotion compared with the traditional femoral cone construct in 5 of the 12 directions measured (P < .05), whereas no statistical difference was found in 7 directions. The new porous titanium metaphyseal tibial cones demonstrated less micromotion in medial varus/valgus (P = .004) and posterior compressive micromotion (P = .002) compared with the traditional porous tantalum system. CONCLUSION: The findings of this biomechanical study demonstrate satisfactory mechanical stability of an anatomical-based porous titanium metaphyseal cone system for femoral and tibial bone loss as measured by micromotion under physiologic loading. The new cone design, in combination with instrumentation that facilitates surgical efficiency, is encouraging. Long-term clinical follow-up is warranted.

来源:J Arthroplasty,查看原文。版权归原作者所有。