无影像导航联合前倾角概念在全髋关节翻修术中的应用
本研究回顾性评估40例(男24例、女16例)无骨水泥全髋关节翻修术,采用无影像导航并按Widmer公式(髋臼杯前倾角+0.7×股骨柄前倾角)定位髋臼杯;平均随访80.7个月,翻修后假体位置无超出安全区者,未观察到脱位或骨溶解,作者认为无影像导航有助于将假体置入安全区。
要点
- 按Widmer公式计算联合前倾角,髋臼杯倾斜40°±10°、联合前倾角37°±10°被视为安全区
- 翻修术后髋臼杯倾斜42.3°±3.1°、前倾25.0°±2.9°、联合前倾36.1°±3.4°,无超出安全区者
- 平均随访80.7个月,未观察到脱位或骨溶解
核心结论
- 作者认为,无影像导航有助于术者在全髋关节翻修术中把假体组件置入安全区;当持续获得该安全区时,这些患者在约6年随访期内未观察到术后脱位。
研究设计与样本
- 回顾性评估40例无骨水泥全髋关节翻修术(男24例、女16例),采用无影像导航联合前倾角概念;平均随访80.7个月。
- 联合前倾角概念用于髋臼杯定位,基于Widmer公式(髋臼杯前倾角+0.7×股骨柄前倾角)。
- 术后髋臼杯倾斜角在标准正位X线片上评估,髋臼杯前倾角与股骨柄前倾角用CT扫描评估。
主要发现
- 翻修股骨柄平均前倾角15.3°±2.9°(范围9.5°-21.5°),保留股骨柄平均前倾角17.4°±9.7°(范围4.2°-29.8°)。
- 翻修术后髋臼杯倾斜42.3°±3.1°(范围32.1°-48.2°)、髋臼杯前倾25.0°±2.9°(范围16.9°-29.5°)、联合前倾36.1°±3.4°(范围27.2°-42.9°)。
- 假体位置相对安全区无超出安全区者,术后未观察到脱位或骨溶解。
关键数值与推荐
- 髋臼杯倾斜40°±10°且基于Widmer公式的髋臼杯与股骨柄联合前倾37°±10°被视为「安全区」。
- 联合前倾角按Widmer公式计算:髋臼杯前倾角+0.7×股骨柄前倾角。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The purpose of the study was to evaluate the implant positions and clinical results of revision total hip arthroplasty (THA) using an imageless navigation with the concept of combined anteversion. METHODS: A total of 40 cementless revision THAs (24 men and 16 women) using an imageless navigation with the concept of combined anteversion were retrospectively evaluated. The concept of combined anteversion was applied in cup positioning based on Widmer's equation (cup anteversion + 0.7 × stem anteversion). The mean follow-up period was 80.7 months. Postoperatively, the inclination of the cup was evaluated on standard anteroposterior view of the radiograph, and the anteversion of the cup and femoral stem was evaluated using computed tomography scan. A cup inclination of 40° ± 10° and combined anteversion of the cup and femoral stem of 37° ± 10° based on Widmer's equation were regarded as the "safe zone." RESULTS: The average anteversion of the revised femoral stems was 15.3° ± 2.9° (range, 9.5°-21.5°), whereas that of the remained femoral stems was 17.4° ± 9.7° (range, 4.2°-29.8°). The inclination, anteversion of the cup, and combined anteversion after revision THA were 42.3° ± 3.1° (range, 32.1°-48.2°), 25.0° ± 2.9° (range, 16.9°-29.5°), and 36.1° ± 3.4° (range, 27.2°-42.9°), respectively. Therefore, the position of the implants, relative to the safe zone, showed no outliers after the revision surgery. Neither dislocation nor osteolysis was observed after the surgery. CONCLUSION: Favorable results of this study indicate that imageless navigation helps the surgeon in placing the components of revision THA in the safe zone. This study also shows that when this safe zone is consistently obtained, then no postoperative dislocations were observed in these patients over the 6-year follow-up period.