文献速递关节外科

金属壳与金属内衬之间的微动腐蚀可能解释R3模块化金属对金属髋关节高失败率

Fretting and Corrosion Between a Metal Shell and Metal Liner May Explain the High Rate of Failure of R3 Modular Metal-on-Metal Hips.
PubMed 文献 · J Arthroplasty · 2016-12-22

对6例取出的带金属内衬的R3髋臼系统进行宏观和微观视觉分析,发现金属内衬背面存在腐蚀,腐蚀区域边界与髋臼壳内表面用于聚乙烯内衬的抗旋转凸片插入位置吻合。扫描电镜显示缝隙腐蚀证据,能量色散X射线分析证实腐蚀碎屑富含钛。作者认为金属内衬的高失败率可能归因于内衬背面腐蚀,后者似乎源于髋臼壳的几何和设计特征。

要点

  • 6例取出的R3髋臼系统金属内衬背面均发现腐蚀。
  • 腐蚀区域边界与髋臼壳内表面用于聚乙烯内衬的抗旋转凸片插入位置吻合。
  • 扫描电镜显示缝隙腐蚀证据,能量色散X射线分析证实腐蚀碎屑富含钛。

核心结论

  • R3髋臼系统金属内衬选项的高失败率可能归因于内衬背面腐蚀,后者似乎源于髋臼壳的几何和设计特征。

研究设计与样本

  • 对6例取出的带金属内衬的R3髋臼系统进行宏观和微观视觉分析。

主要发现

  • 视觉分析发现金属内衬背面存在腐蚀,腐蚀区域边界与髋臼壳内表面用于聚乙烯内衬的抗旋转凸片插入位置吻合。
  • 扫描电镜显示缝隙腐蚀证据,能量色散X射线分析证实腐蚀碎屑富含钛。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The R3 acetabular system used with its metal liner has higher revision rates when compared to its ceramic and polyethylene liner. In June 2012, the medical and healthcare products regulatory agency issued an alert regarding the metal liner of the R3 acetabular system. METHODS: Six retrieved R3 acetabular systems with metal liners underwent detailed visual analysis using macroscopic and microscopic techniques. RESULTS: Visual analysis discovered corrosion on the backside of the metal liners. There was a distinct border to the areas of corrosion that conformed to antirotation tab insertions on the inner surface of the acetabular shell, which are for the polyethylene liner. Scanning electron microscopy indicated evidence of crevice corrosion, and energy-dispersive X-ray analysis confirmed corrosion debris rich in titanium. CONCLUSION: The high failure rate of the metal liner option of the R3 acetabular system may be attributed to corrosion on the backside of the liner which appear to result from geometry and design characteristics of the acetabular shell.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Ilo KC, Derby EJ, Whittaker RK, et al. Fretting and Corrosion Between a Metal Shell and Metal Liner May Explain the High Rate of Failure of R3 Modular Metal-on-Metal Hips. J Arthroplasty. 2016. PMID: 28159422. doi:10.1016/j.arth.2016.12.024