全髋关节置换术中扩髓器手柄设计改变股骨受力分布

Broach Handle Design Changes Force Distribution in the Femur During Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2016-12-29

研究采用六轴测力传感器实验模型和有限元分析,评估4种不同偏移和曲度的扩髓器手柄在徒手敲击时传向骨内的力学比率。高偏移手柄平均多损失4%锤击力至水平轴;有限元分析中骨质疏松骨内水平力从4%放大至最大52%;两种方法均证实大偏移手柄使力矩-力比增至163%-235%。

要点

  • 高偏移手柄平均多损失4%锤击力至水平轴。
  • 有限元分析中骨质疏松骨内水平向皮质骨的力从4%放大至最大52%。
  • 实验模型和有限元分析均证实大偏移手柄使力矩-力比增至163%-235%,使扩髓器近远端向皮质骨旋转。

核心结论

  • 扩髓器手柄设计是全髋关节置换术中传至扩髓器及最终传至骨的合力重要决定因素。
  • 不需要的离轴力和增强的旋转动力学可能在股骨髓腔准备中参与术中骨折的发生。

研究设计与样本

  • 实验模型使用六轴测力传感器测量4种不同扩髓器手柄产生的力分布,每种手柄具有递增的偏移和曲度。
  • 手柄分别接受敲击并评估动态变量,随后用高分辨率光学系统数字化并行有限元分析,考虑松质骨并改变锤击位置。
  • 计算离轴力、扩髓构造力矩和周围骨内应力。

主要发现

  • 实验模型中,高偏移手柄平均多损失4%锤击力至水平轴。
  • 有限元分析采用弹性模量估算骨质疏松松质骨内扩髓时,水平向皮质骨方向的力从4%放大至最大52%。
  • 实验构造和有限元分析均证实较大偏移手柄使力矩-力比增至163%-235%,使扩髓器近端和远端向皮质骨旋转。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Curved broach handles were developed to overcome limited surgical exposures during total hip arthroplasty. Some authors report increased intraoperative fracture rates during limited exposures. This study evaluates mechanical force ratios transmitted to the bone while broaching with curved vs straight handles. METHODS: An experimental model utilized a 6-axis load cell to measure force distributions produced by 4 different broach handles, each with increasing offset and curvature. Handles were separately impacted and dynamic variables assessed. Handles were then digitized using a high-resolution optical system and a finite element analysis (FEA) was performed to account for trabecular bone and vary the location of mallet impact. Off-axis forces, broaching construct moments, and stress within surrounding bone were computed. RESULTS: Using the experimental model, high-offset handles lost on average 4% more hammering force to the horizontal axis. When the FEA utilized moduli of elasticity to estimate broaching through osteoporotic trabecular bone, horizontally displaced forces (toward cortical bone) were magnified from 4% to a maximum value of 52%. Both the experimental construct and FEA confirmed that larger offset handles increase moment-to-force ratios up to 163%-235%, thus rotating the proximal and distal ends of the broach toward cortical bone. CONCLUSION: Broach handle design is an important determinant of resultant forces transmitted to the broach (and ultimately the bone) during total hip arthroplasty. Unwanted off-axis forces and enhanced rotational dynamics may play a role in intraoperative fractures during femoral canal preparation.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Greenhill DA, Abbasi P, Darvish K, et al. Broach Handle Design Changes Force Distribution in the Femur During Total Hip Arthroplasty. J Arthroplasty. 2016. PMID: 28108173. doi:10.1016/j.arth.2016.12.031

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。