应用同步双平面采集测量髋臼位置对聚乙烯内衬磨损的影响
研究纳入43髋金属对高交联聚乙烯全髋关节置换术后至少10年随访患者,采用改良放射立体分析测量髋臼外倾角和前倾角与线性磨损率的关系。平均随访12.3±1.2年,平均线性磨损率为0.066±0.066 mm/y,外倾角(P=.82)和前倾角(P=.11)均与聚乙烯磨损率无相关性。结论认为高交联聚乙烯长期磨损率极低,且该特性与髋臼位置无关。
要点
- 高交联聚乙烯全髋关节置换术后长期平均线性磨损率为0.066±0.066 mm/y。
- 髋臼外倾角与聚乙烯磨损率无相关性(P=.82)。
- 髋臼前倾角与聚乙烯磨损率无相关性(P=.11)。
核心结论
- 在高交联聚乙烯全髋关节置换术后超过10年的长期随访中,高交联聚乙烯磨损率极低,且该优异摩擦学性能与髋臼位置无关。
- 低磨损率凸显了金属对高交联聚乙烯的优异结果,并支持其在全髋关节置换术中的应用。
研究设计与样本
- 利用机构关节置换数据库识别使用相同部件设计的金属对高交联聚乙烯初次全髋关节置换术患者,最低随访10年。
- 采用改良放射立体分析检查设置,在单次立体放射立体分析采集中重建标准正位和交叉侧位检查。
- 使用同一组X线片测量髋臼外倾角和前倾角。
- 共纳入43髋进行分析,平均随访12.3±1.2年。
主要发现
- 平均线性磨损率计算为0.066±0.066 mm/y。
- 髋臼外倾角与聚乙烯磨损率无相关性(P=.82)。
- 髋臼前倾角与聚乙烯磨损率无相关性(P=.11)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Studies that have previously examined the relationship between acetabular component inclination angle and polyethylene wear have shown increased wear of conventional polyethylene with high inclination angles. To date, there are no long-term in vivo studies examining the correlation between cup position and polyethylene wear with highly crosslinked polyethylene. METHODS: An institutional arthroplasty database was used to identify patients who had metal-on-highly crosslinked polyethylene primary total hip arthroplasty using the same component design with a minimum follow-up of 10 years. A modified radiostereometric analysis examination setup was utilized, recreating standard anteroposterior and cross-table lateral examinations in a single stereo radiostereometric analysis acquisition. The same radiographs were used to measure inclination angle and anteversion. RESULTS: A total of 43 hips were included for analysis in this study. Average follow-up was 12.3 ± 1.2 years. The average linear wear rate was calculated to be 0.066 ± 0.066 mm/y. Inclination angle was not correlated with polyethylene wear rate (P = .82). Anteversion was also not correlated with polyethylene wear rate (P = .11). CONCLUSION: At long-term follow-up of >10 years, highly crosslinked polyethylene has a very low wear rate. This excellent tribology is independent of acetabular position. The low wear rate highlights the excellent results of metal on highly crosslinked polyethylene, and supports its use in total hip arthroplasty.
引用
- Goyal P, Howard JL, Yuan X, et al. Effect of Acetabular Position on Polyethylene Liner Wear Measured Using Simultaneous Biplanar Acquisition. J Arthroplasty. 2016. PMID: 28087161. doi:10.1016/j.arth.2016.11.057
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。