全髋关节置换术使用双动髋臼假体后的脱位:一项系统综述
该系统综述纳入59篇文献共17908例全髋关节置换术,评估双动髋臼假体在初次和翻修全髋关节置换术后的脱位率。初次置换组平均脱位率为0.9%,翻修组为3.0%;假体内脱位率分别为0.7%和1.3%。作者认为双动髋臼假体可有效降低不稳定风险,但需权衡额外模块化和假体内脱位这一新失败模式。
要点
- 初次全髋关节置换术使用双动髋臼假体的平均脱位率为0.9%。
- 翻修全髋关节置换术使用双动髋臼假体的平均脱位率为3.0%。
- 假体内脱位率在初次置换为0.7%,翻修置换为1.3%。
核心结论
- 基于现有数据,双动髋臼假体在初次和翻修全髋关节置换术后均能有效降低不稳定风险。
- 这一获益需与对额外模块化和假体内脱位这一新失败模式的持续担忧相权衡。
- 需要更长期的研究来评估这些较新材料与既往几代材料相比的功能表现。
研究设计与样本
- 按照系统综述和Meta分析首选报告条目(PRISMA)指南进行的系统综述。
- 检索Pubmed/Medline、Cochrane Library和Embase(Scopus)中1974年1月至2016年3月间的英文文献,检索词包括「dual mobility」「dual-mobility」「tripolar」「double-mobility」「double mobility」「hip」「cup」「socket」等不同组合。
- 共59篇文章符合纳入标准,包含17908例全髋关节置换术,分为初次全髋关节置换术组和翻修全髋关节置换术组。
- 由两名研究者独立提取人口学资料、手术为初次还是翻修、随访时间、假体设计、股骨头直径及髋臼假体固定方式等数据。
主要发现
- 初次全髋关节置换术组平均脱位率为0.9%,翻修全髋关节置换术组平均脱位率为3.0%。
- 初次全髋关节置换术组平均假体内脱位率为0.7%,翻修全髋关节置换术组平均假体内脱位率为1.3%。
局限性
- 需要更长期的研究来评估这些较新材料与既往几代材料相比的功能表现。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: The aim of this systematic review was to report the rate of dislocation following the use of dual mobility (DM) acetabular components in primary and revision total hip arthroplasty (THA). MATERIALS AND METHODS: A systematic review of the literature according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses guidelines was performed. A comprehensive search of Pubmed/Medline, Cochrane Library and Embase (Scopus) was conducted for English articles between January 1974 and March 2016 using various combinations of the keywords "dual mobility", "dual-mobility", "tripolar", "double-mobility", "double mobility", "hip", "cup", "socket". The following data were extracted by two investigators independently: demographics, whether the operation was a primary or revision THA, length of follow-up, the design of the components, diameter of the femoral head, and type of fixation of the acetabular component. RESULTS: In all, 59 articles met our inclusion criteria. These included a total of 17 908 THAs which were divided into two groups: studies dealing with DM components in primary THA and those dealing with these components in revision THA. The mean rate of dislocation was 0.9% in the primary THA group, and 3.0% in the revision THA group. The mean rate of intraprosthetic dislocation was 0.7% in primary and 1.3% in revision THAs. CONCLUSION: Based on the current data, the use of DM acetabular components are effective in minimising the risk of instability after both primary and revision THA. This benefit must be balanced against continuing concerns about the additional modularity, and the new mode of failure of intraprosthetic dislocation. Longer term studies are needed to assess the function of these newer materials compared with previous generations. Cite this article: Bone Joint J 2017;99-B(1 Supple A):18-24.