假体周围股骨骨折及其规避:股骨假体设计对假体周围股骨骨折风险增加的贡献
该系统综述检索多个期刊数据库及美国骨科医师学会近期会议录,纳入34篇提供足够假体描述的文章,汇总342 719例初次全髋关节置换中的1691例假体周围股骨骨折。非骨水泥股骨假体与更高的假体周围股骨骨折数量显著相关(p < 0.001);单楔形和双楔形(fit-and-fill)假体的骨折率较解剖型、全涂层和锥形/圆形柄增加三倍(p < 0.001);骨水泥柄中载荷锥形(Exeter)柄的骨折多于复合梁(Charnley)柄(p = 0.004)。
要点
- 非骨水泥股骨假体与更高的假体周围股骨骨折数量显著相关(p < 0.001)。
- 单楔形和双楔形(fit-and-fill)假体的骨折率较解剖型、全涂层和锥形/圆形柄增加三倍(p < 0.001)。
- 骨水泥柄中载荷锥形(Exeter)柄的骨折多于复合梁(Charnley)柄(p = 0.004)。
核心结论
- 假体周围股骨骨折文献中提供详细假体信息的研究很少。
- 非骨水泥假体尤其是单楔形和双楔形假体的假体周围股骨骨折率最高,多数研究建议不要在老年骨质疏松患者中使用。
- 本综述说明注册登记和未来假体周围股骨骨折研究需要记录假体名称和信息以供未来分析。
- 未来需要比较现代假体的生物力学研究来明确假体设计对假体周围股骨骨折风险的精确贡献。
研究设计与样本
- 对多个期刊数据库和美国骨科医师学会近期会议录进行系统检索,纳入提供足够假体描述的临床文章,由两名评价者审阅。
- 对假体载荷至失效的基础研究进行综述,以识别临床与基础文献中的相似结论。
- 初始识别596篇文章,其中34篇符合分析条件。
主要发现
- 汇总342 719例初次全髋关节置换中的1691例假体周围股骨骨折,非骨水泥股骨假体的假体周围股骨骨折数量显著更高(p < 0.001)。
- 单楔形和双楔形(fit-and-fill)股骨假体与解剖型、全涂层和锥形/圆形柄相比,假体周围股骨骨折率增加三倍(p < 0.001)。
- 骨水泥柄中,载荷锥形(Exeter)柄与复合梁(Charnley)柄相比假体周围股骨骨折更多(p = 0.004)。
- 基础文献中比较非骨水泥假体设计的研究非常少。
关键数值与推荐
- 临床文章需提供足够假体描述方纳入分析。
- 非骨水泥股骨假体的假体周围股骨骨折数量显著高于对照(p < 0.001)。
- 单楔形和双楔形(fit-and-fill)股骨假体的假体周围股骨骨折率较解剖型、全涂层和锥形/圆形柄增加三倍(p < 0.001)。
- 骨水泥载荷锥形(Exeter)柄的假体周围股骨骨折多于复合梁(Charnley)柄(p = 0.004)。
临床怎么用
- 多数研究建议不要在老年骨质疏松患者中使用非骨水泥单楔形和双楔形假体。
- 本综述说明注册登记和未来假体周围股骨骨折研究应记录假体名称和信息以供未来分析。
- 未来需要比较现代假体的生物力学研究来明确假体设计对假体周围股骨骨折风险的精确贡献。
局限性
- 假体周围股骨骨折文献中提供详细假体信息的研究很少。
- 基础文献中比较非骨水泥假体设计的研究非常少。
- 未来需要比较现代假体的生物力学研究来明确假体设计对假体周围股骨骨折风险的精确贡献。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used. PATIENTS AND METHODS: A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature. RESULTS: In total 596 articles were initially identified, with 34 being eligible for analysis. Aggregate analysis of 1691 PFFs in 342 719 primary THAs revealed a significantly higher number of PFFs with cementless femoral implants (p < 0.001). Single-wedge and double-wedge (fit-and-fill) femoral implants were associated with a threefold increase in PFF rates (p < 0.001) compared with anatomical, fully coated and tapered/rounded stems. Within cemented stems, loaded-taper (Exeter) stems were associated with more PFFs than composite-beam (Charnley) stems (p = 0.004). Review of the fundamental literature revealed very few studies comparing cementless component designs. CONCLUSION: Very few studies within the PFF literature provide detailed implant information. Cementless implants, specifically those of single-wedge and double-wedge, have the highest PFF rates in the literature, with most investigations recommending against their use in older patients with osteoporotic bone. This review illustrates the need for registries and future PFF studies to record implant name and information for future analysis. Furthermore, future biomechanical investigations comparing modern implants are needed to clarify the precise contribution of implant design to PFF risk. Cite this article: Bone Joint J 2017;99-B(1 Supple A):50-9.