髋臼金属壳内锥角对陶瓷内衬安装不良及分离力的影响
研究由三名外科医生手动将陶瓷内衬插入两种不同锥角(10°和18°)的金属壳,比较陶瓷内衬的安装不良率和推出强度。结果显示10°金属壳的安装不良率高于18°金属壳(23.3% vs 0%,P < .05),而10°锥度杯的平均分离力高于18°锥度杯(1148.8 ± 46.7 N vs 389.7 ± 108.3 N,P < .01)。作者建议外科医生在使用内锥角为10°的多轴承金属壳时对陶瓷内衬的安装不良保持谨慎。
要点
- 10°金属壳的陶瓷内衬安装不良率高于18°金属壳(23.3% vs 0%,P < .05)。
- 10°锥度杯的平均分离力高于18°锥度杯(1148.8 ± 46.7 N vs 389.7 ± 108.3 N,P < .01)。
- 作者建议使用内锥角10°多轴承金属壳时谨慎对待陶瓷内衬安装不良。
核心结论
- 作者认为,外科医生在使用内锥角为10°的多轴承金属壳时应谨慎对待陶瓷内衬的安装不良。
研究设计与样本
- 三名外科医生手动将陶瓷内衬插入两种不同锥角(10°和18°)的金属壳设计中,比较陶瓷内衬的安装不良率和推出强度(即陶瓷内衬从金属壳分离的力)。
主要发现
- 10°金属壳的陶瓷内衬安装不良率为23.3%,18°金属壳为0%(P < .05)。
- 10°锥度杯的平均分离力为1148.8 ± 46.7 N,高于18°锥度杯的389.7 ± 108.3 N(P < .01)。
临床怎么用
- 作者建议,当外科医生使用内锥角为10°的多轴承金属壳时,应谨慎对待陶瓷内衬的安装不良。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Malseating of ceramic liner appeared as a matter of concern in multibearing metal, although multibearing cup can be coupled with hard liners as well as polyethylene liner. In this metal shell, the inner taper angle should be 10° for the modularity, while standard metal shells for ceramic liner have an inner taper angle of 18°. However, there has been no study in the effect of taper angle to the risk of malseating. We evaluated whether the taper angle of metal shell might affect the malseating of ceramic liner, and dissociation force of ceramic liner from metal shell. METHODS: Three surgeons manually inserted ceramic liners into two designs of metal shell with different tapers angles (10° and 18°). We compared malseating rates of ceramic liners and push-out strengths, which means dissociation force of the ceramic liner from the metal shell, between these two metal shell designs. RESULTS: The malseating rates in 10° metal shell were higher than those in 18° metal shell (23.3% vs 0%, P < .05). The mean dissociation force (1148.8 ± 46.7 N) in 10° taper cup was higher than that (389.7 ± 108.3 N) in 18° taper cup (P < .01). CONCLUSION: Our results suggest that surgeon should be cautious about malseating of ceramic liner when using multi-bearing metal shell with inner taper angle of 10°. CLINICAL RELEVANCE: When surgeon use multi-bearing metal shell with inner taper angle of 10°, our results suggest that surgeon should be cautious about malseating of ceramic liner.
引用
- Lee YK, Kim KC, Jo WL, et al. Effect of Inner Taper Angle of Acetabular Metal Shell on the Malseating and Dissociation Force of Ceramic Liner. J Arthroplasty. 2016. PMID: 28007373. doi:10.1016/j.arth.2016.11.002