定制型与患者尺寸型后交叉韧带保留型全膝关节置换术:基于移动透视的体内运动学研究
研究采用移动透视对比24例定制型(CIM)后交叉韧带保留型全膝关节置换术(TKA)与14例非对称髁后交叉韧带保留型TKA受试者的体内运动学。CIM TKA在负重屈膝角度、股骨后滚、轴向旋转方面表现更优, paradoxical滑动与反向旋转发生极少,早期至中期屈曲稳定性更好,运动学更接近正常膝关节。
要点
- CIM TKA负重屈膝角度大于传统后交叉韧带保留型TKA。
- CIM TKA轴向旋转显著大于传统PCR TKA(P=0.05)。
- 仅CIM TKA患者在完全伸直时出现股骨内旋,与传统PCR TKA相比 paradoxical滑动和反向旋转极少。
核心结论
- CIM TKA表现出更接近正常膝关节的运动学,与传统PCR TKA相比具有更正常运动的优势。
研究设计与样本
- 使用移动透视分析24例定制型(CIM)后交叉韧带保留型TKA受试者和14例非对称髁后交叉韧带保留型TKA受试者的胫股运动学,受试者完成负重深屈膝和从坐位站起动作。
主要发现
- CIM TKA受试者负重屈膝角度大于传统后交叉韧带保留型(PCR)TKA设计。
- 屈膝过程中CIM TKA受试者股骨后滚持续多于传统PCR TKA受试者。
- CIM TKA轴向旋转显著大于传统PCR TKA(P=0.05)。
- 仅CIM TKA患者在完全伸直时出现股骨内旋,与传统PCR TKA相比 paradoxical滑动和反向旋转极少。
- CIM TKA受试者早期至中期屈曲抬离极小,稳定性优于传统PCR TKA受试者。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Historically, knee arthroplasties have been designed using average patient anatomy. Recent advances in imaging and manufacturing have facilitated the development of customized prostheses designed to fit the unique shape of individual patients. The purpose of this study is to determine if improving implant design through customized total knee arthroplasty (TKA) improves kinematic function. METHODS: Using state-of-the-art mobile fluoroscopy, tibiofemoral kinematics were analyzed for 24 subjects with a customized individually made (CIM), cruciate-retaining TKA, and 14 subjects having an asymmetric condylar cruciate-retaining TKA. Subjects performed a weight-bearing deep knee bend and a rise from a seated position. Each patient was evaluated for weight-bearing range of motion, femorotibial translation, femorotibial axial rotation, and condylar liftoff occurrence. RESULTS: Subjects having a CIM TKA experienced greater weight-bearing knee flexion compared with the traditional posterior cruciate-retaining (PCR) TKA design. During flexion, the CIM TKA subjects consistently exhibited more posterior femoral rollback than the traditional PCR TKA subjects. The CIM TKA was found to have statistically greater axial rotation compared with the traditional PCR TKA (P = .05). Of note, only the CIM TKA patients experienced femoral internal rotation at full extension, as exhibited in a normal knee. Compared with the traditional PCR TKA, the CIM TKAs demonstrated minimal occurrences of paradoxical sliding and reverse rotation during flexion and extension. The CIM TKA subjects showed minimal liftoff and hence better stability in earlyflexion to midflexion compared with the traditional PCR subjects. CONCLUSION: The CIM TKA demonstrated kinematics more similar to a normal knee. Therefore, using customized implant technology through CIM TKA designs affords benefits including more normal motion compared with a traditional PCR TKA.
引用
- Zeller IM, Sharma A, Kurtz WB, et al. Customized versus Patient-Sized Cruciate-Retaining Total Knee Arthroplasty: An In Vivo Kinematics Study Using Mobile Fluoroscopy. J Arthroplasty. 2016. PMID: 27814916. doi:10.1016/j.arth.2016.09.034
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。