通用埃克塞特股骨柄假体骨折已报告病例的分析
该研究汇总了60例通用埃克塞特股骨柄假体骨折(32例柄骨折、28例颈骨折)的临床、手术、影像及取回数据,分析患者相关、手术相关及解剖/假体相关危险因素。颈骨折多与患者因素(体重增加、活动量大)和假体因素(使用加长股骨头)相关;柄骨折多与可纠正的手术因素相关,主要为柄尺寸过小或内侧支撑不足(84%)。
要点
- 颈骨折多与患者因素(体重增加、活动量大)及加长股骨头相关。
- 柄骨折多与可纠正的手术因素相关,主要为柄尺寸过小或内侧支撑不足(84%)。
- 柄骨折更常见于较小尺寸(35.5至44#1),占84%。
核心结论
- 颈骨折最常与患者相关因素(体重增加和活动量大)及假体相关因素(使用加长股骨头)相关。
- 柄骨折最常与可纠正的手术相关原因相关,主要为柄尺寸过小或内侧支撑不足(84%)。
研究设计与样本
- 研究汇总了临床、手术、放射学及取回数据。
- 数据来自60例患者,其中32例柄骨折、28例颈骨折。
主要发现
- 颈骨折与柄骨折患者的骨折时平均年龄相似(分别为69岁和67岁)。
- 颈骨折中77%发生于男性,柄骨折中52%发生于男性。
- 颈骨折平均体重为107 kg,柄骨折为96.5 kg;颈骨折中68%为肥胖或病态肥胖,柄骨折中38%为肥胖或病态肥胖。
- 颈骨折平均至骨折时间为78个月(范围36-144个月),柄骨折为76个月(范围2-155个月)。
- 44#2和44#3是与颈骨折相关的最常见尺寸;柄骨折更常见(84%)于较小尺寸(35.5至44#1)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Between 1991 and 2008, approximately 80 cases of fracture (neck or stem) have been reported. This study aimed at determining factors predisposing to implant fracture. METHODS: Clinical, surgical, radiological, and retrieval data were collated. Risk factors associated with fracture were categorized to patient related (weight and activity levels), surgical related (poor medial support, component size, and placement), and anatomic/implant related (head size/offset). RESULTS: Data was available on 60 patients (32 stem and 28 neck fractures). Mean patient age at fracture was similar for both neck and stem fractures (69 years, 67 years, respectively). Also, 77% neck and 52% stem fractures occurred in men. Mean weight was 107 kg in neck and 96.5 kg in stem fractures with 68% neck and 38% stem fractures either obese or morbidly obese. Mean time to fracture was 78 months (range, 36-144 months) for neck and 76 months (range, 2-155 months) for stem fractures. 44#2 and 44#3 were the most common sizes associated with neck fractures. Stem fractures occurred more commonly (84%) in the smaller sizes (35.5 to 44#1). Elongated femoral heads were used in 69% neck and 14% stem fractures. CONCLUSION: Neck fractures were most commonly associated with patient-related (increased weight and activity) and implant-related (use of an elongated femoral head) factors. Stem fractures were most commonly associated with correctable surgical-related causes, predominantly secondary to stem undersizing or inadequate medial support (84%).
引用
- Bolland BJ, Wilson MJ, Howell JR, et al. An Analysis of Reported Cases of Fracture of the Universal Exeter Femoral Stem Prosthesis. J Arthroplasty. 2016. PMID: 27843041. doi:10.1016/j.arth.2016.09.032
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。