十年随访:传统聚乙烯与顺序退火高交联聚乙烯衬垫在非骨水泥全膝置换中的磨损与假体迁移随机对照试验

Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis.
PubMed 文献 · Bone Joint J · 2026-10-01

这项双盲随机对照试验纳入96例非骨水泥Triathlon固定平台后稳定型全膝置换,采用模型放射立体测量分析比较传统聚乙烯与顺序退火高交联聚乙烯衬垫的十年磨损及胫骨、股骨假体迁移。高交联聚乙烯组五年后内侧磨损无进展(0.15 mm至0.13 mm),传统聚乙烯组磨损率增加(0.22 mm至0.46 mm);两组假体迁移和患者报告结局在十年时无临床相关差异。

要点

  • 高交联聚乙烯衬垫五年后磨损无进展,传统聚乙烯衬垫磨损持续增加。
  • 十年时两组胫骨和股骨假体迁移及患者报告结局无临床相关差异。
  • 非骨水泥全膝置换中使用高交联聚乙烯衬垫在磨损和假体迁移方面是安全的。

核心结论

  • 高交联聚乙烯组五年后未见明显增加的磨损,而传统聚乙烯组磨损持续存在。
  • 非骨水泥全膝置换假体在初始迁移后于十年随访时保持稳定。
  • 就磨损和假体迁移而言,非骨水泥全膝置换中使用高交联聚乙烯衬垫是安全的。

研究设计与样本

  • 双盲随机对照试验纳入96例非骨水泥Triathlon固定平台后稳定型全膝置换。
  • 采用模型放射立体测量分析测量衬垫磨损及胫骨、股骨假体迁移至术后十年。
  • 术前及随访时收集患者报告结局。

主要发现

  • 高交联聚乙烯组内侧磨损从五年0.15 mm至十年0.13 mm,五年后无进展。
  • 传统聚乙烯组磨损率从五年0.22 mm增加至十年0.46 mm。
  • 十年时两组胫骨和股骨假体迁移及患者报告结局无临床相关差异。

关键数值与推荐

  • 高交联聚乙烯组十年胫骨假体最大总点运动为0.99 mm(95%CI 0.83至1.18)。
  • 传统聚乙烯组十年胫骨假体最大总点运动为1.05 mm(95%CI 0.87至1.24)。
  • 高交联聚乙烯组十年股骨假体最大总点运动为1.16 mm(95%CI 0.99至1.35)。
  • 传统聚乙烯组十年股骨假体最大总点运动为1.04 mm(95%CI 0.87至1.22)。

临床怎么用

  • 非骨水泥全膝置换可选用高交联聚乙烯衬垫,十年磨损和假体迁移结果安全。
  • 传统聚乙烯衬垫五年后磨损持续增加,需关注长期磨损相关风险。
  • 非骨水泥假体初始迁移后十年内保持稳定,支持其长期固定可靠性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The aims of this randomized controlled trial (RCT) was to compare the long-term insert wear and migration of the tibial and femoral components between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). METHODS: A double-blinded RCT was conducted including 96 TKAs. The wear of the insert, and migration of both the tibial and femoral components, was measured with model-based radiostereometric analysis up to ten years postoperatively. Patient-reported outcome measures (PROMs) were collected preoperatively and at follow-up. RESULTS: There was no progression of (medial) wear (from 0.15 mm at five years to 0.13 mm at ten years) for the HXLPE group after five years, whereas the wear rate increased for the PE group (from 0.22 mm at five years to 0.46 mm at ten years). There was no clinically relevant difference in terms of migration of the tibial and femoral component or PROMs between the HXLPE and PE groups at ten years. Migration of the tibial component (maximal total point motion; MTPM) after the ten-year follow-up for the HXLPE group was 0.99 mm (95% CI 0.83 to 1.18), and for the PE group, was 1.05 mm (95% CI 0.87 to 1.24). Migration of the femoral component (MTPM) after the ten-year follow-up for the HXLPE group was 1.16 mm (95% CI 0.99 to 1.35), and for the PE group, MTPM was 1.04 mm (95% CI 0.87 to 1.22). CONCLUSION: There was no discernible increased wear for the HXLPE group after five years, whereas the wear for the PE group continued. After the initial migration, the uncemented TKA components remain stable at the ten-year follow-up. The use of HXLPE inserts are safe, when used in uncemented TKA, when considering wear and component migration.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. de Ridder R, Kaptein BL, Pijls BG, et al. Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis. Bone Joint J. 2026. PMID: 42853089. doi:10.1302/0301-620X.108B10.BJJ-2026-0426.R1