文献速递关节外科运动医学

全髋关节置换术后挥拍运动中的在体三维髋关节运动学:动态活动度与内衬-颈接触评估

In vivo 3D hip kinematics during racket swinging after total hip arthroplasty : evaluation of dynamic range of motion and liner-to-neck contact.
PubMed 文献 · Bone Joint J · 2026-10-01

本研究对19例(24髋)全髋置换术后进行挥拍运动的患者,采用平板探测器和模型图像配准技术分析正手与反手挥拍时的在体三维髋关节运动学。结果显示挥拍时平均总轴向旋转为32°,2髋(8%)在高边内衬最大外旋时发生内衬-颈接触,9髋(20%)后间隙小于10°。高边内衬、较大髋臼假体前倾角和最大外旋时较大髋伸展角与后间隙减小相关。

要点

  • 全髋置换术后挥拍运动在运动学安全范围内进行,平均总轴向旋转32°。
  • 内衬-颈接触发生率低(2/24髋,8%),均发生于高边内衬最大外旋时。
  • 高边内衬、较大髋臼前倾角和最大外旋时较大髋伸展角与后间隙减小相关。

核心结论

  • 全髋关节置换术后在体挥拍运动在运动学安全范围内进行,无过度轴向旋转。
  • 内衬-颈接触不常见,但高边内衬和过大髋臼假体前倾角与最大外旋时后间隙减小相关。
  • 挥拍运动总体可行,但需谨慎选择内衬和假体方位以降低内衬-颈接触风险。

研究设计与样本

  • 研究纳入19例(24髋)初次全髋关节置换术后进行挥拍运动的患者。
  • 采用平板探测器获取正手和反手挥拍动态放射学图像,并用模型图像配准技术分析髋关节运动学。
  • 利用术后CT数据和运动学分析评估内衬-颈接触与间隙,多变量回归分析确定后间隙减小的相关因素。

主要发现

  • 挥拍时平均总轴向髋旋转为32°,髋关节在整个运动中保持轻度屈曲和外展。
  • 24髋中2髋(8%)发生内衬-颈接触,均位于高边内衬最大外旋时,未观察到骨-骨接触。
  • 9髋(20%)在伸展和外旋时后间隙小于10°,提示后侧接触风险增加。
  • 多变量回归分析显示高边内衬使用、较大髋臼假体前倾角和最大外旋时较大髋伸展角与后间隙减小显著相关。

关键数值与推荐

  • 后间隙小于10°被视为后侧接触风险增加,见于9髋(20%)。
  • 内衬-颈接触发生于2髋(8%),均在高边内衬最大外旋时。

临床怎么用

  • 全髋关节置换术后挥拍运动总体可行,但应谨慎选择内衬类型和假体方位。
  • 高边内衬和较大髋臼前倾角可能增加最大外旋时后侧内衬-颈接触风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Interest in sports participation after total hip arthroplasty (THA) has grown, and some patients return to playing racket sports after THA. However, dynamic hip kinematics and the risk of liner-to-neck contact during racket swinging remain unclear. This study aimed to evaluate in vivo, 3D dynamic hip kinematics during racket swinging after THA, and to identify factors associated with reduced liner-to-neck clearance, which may indicate an increased risk of liner-to-neck contact. METHODS: A total of 19 patients (24 hips) who were playing racket sports after primary THA were evaluated. Dynamic radiological images of forehand and backhand swings were acquired using a flat panel detector. Hip joint kinematics were analyzed using a model-image registration technique. Liner-to-neck contact and clearance were assessed using postoperative CT data and kinematic analysis. Multivariable regression analyses identified factors associated with reduced posterior clearance. RESULTS: The mean total axial hip rotation during racket swinging was 32°, with the hip maintained in slight flexion and abduction throughout the motion. Liner-to-neck contact occurred in two of 24 hips (8%), all at the elevated rim during maximum external rotation (ER). No bone-to-bone contact was observed. Posterior clearance below 10°, indicating an increased risk of posterior contact, was observed in nine hips (20%) during both extension and external rotation. Multivariable regression analysis identified elevated liner use, greater acetabular component anteversion, and greater hip extension angle at maximum ER as significant factors associated with reduced posterior clearance. CONCLUSION: In vivo racket swinging after THA was performed within a kinematically safe range of motion without excessive axial rotation. Although liner-to-neck contact was infrequent, elevated liners and excessive acetabular component anteversion were associated with reduced posterior clearance during maximum external rotation. While racket swinging appears generally permissible after THA, careful liner selection and implant orientation are important to minimize the risk of liner-to-neck contact.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Nakao Y, Hamai S, Murakami K, et al. In vivo 3D hip kinematics during racket swinging after total hip arthroplasty : evaluation of dynamic range of motion and liner-to-neck contact. Bone Joint J. 2026. PMID: 42853108. doi:10.1302/0301-620X.108B10.BJJ-2026-0030.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。