文献速递关节外科创伤骨科

老年股骨颈骨折:非骨水泥干骺端填充带领股骨柄假体周围骨折风险高于骨水泥固定

Increased Risk of Periprosthetic Femoral Fracture With Cementless Collared Metadiaphyseal-Filling Stems Compared to Cemented Fixation for Femoral Neck Fractures in Patients Aged 65 Years or Older: An American Joint Replacement Registry Analysis.
PubMed 文献 · J Arthroplasty · 2026-08-19

基于美国关节置换登记系统2012至2022年65岁及以上股骨颈骨折行关节置换的患者数据,比较非骨水泥干骺端填充带领(CMDF)股骨柄与骨水泥柄的假体周围骨折(PPFx)风险。结果显示CMDF柄因PPFx行翻修或切开复位内固定的风险显著高于骨水泥柄(aHR 1.7,95%CI 1.1-2.7,P=0.02),但因无菌性松动行翻修的风险显著低于骨水泥柄(aHR 0.5,95%CI 0.2-0.9,P=0.04)。

要点

  • 65岁及以上股骨颈骨折关节置换,CMDF非骨水泥柄PPFx翻修风险高于骨水泥柄(aHR 1.7)。
  • CMDF非骨水泥柄因无菌性松动翻修风险低于骨水泥柄(aHR 0.5)。
  • 选择固定方式需权衡PPFx风险与无菌性松动风险。

核心结论

  • 在65岁及以上股骨颈骨折关节置换中,非骨水泥CMDF柄较骨水泥柄与更高的PPFx翻修风险相关(aHR 1.7,95%CI 1.1-2.7,P=0.02)。
  • 非骨水泥CMDF柄因无菌性松动行翻修的风险显著低于骨水泥柄(aHR 0.5,95%CI 0.2-0.9,P=0.04)。
  • 该发现支持骨水泥柄用于股骨颈骨折治疗,但需权衡其较高的无菌性松动风险。

研究设计与样本

  • 美国关节置换登记系统(AJRR)2012至2022年数据,链接美国医疗保险和医疗补助服务中心(CMS)数据。
  • 纳入65岁及以上股骨颈骨折行半髋或全髋关节置换、接受非骨水泥CMDF柄或骨水泥柄的患者。
  • 采用病因特异性Cox比例风险模型评估假体设计与翻修结局的关联,并校正潜在混杂因素。

主要发现

  • 非骨水泥CMDF柄因PPFx行翻修或切开复位内固定的风险高于骨水泥柄(aHR 1.7,95%CI 1.1-2.7,P=0.02)。
  • 非骨水泥CMDF柄因无菌性松动行翻修的风险显著低于骨水泥柄(aHR 0.5,95%CI 0.2-0.9,P=0.04)。
  • 其他次要结局的校正分析未发现显著差异。

临床怎么用

  • 对65岁及以上股骨颈骨折行关节置换时,应认识到非骨水泥CMDF柄的PPFx翻修风险高于骨水泥柄。
  • 选择骨水泥柄时需考虑其较高的无菌性松动翻修风险,个体化权衡固定方式。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Cemented femoral fixation is recommended when treating a femoral neck fracture with arthroplasty. Prior studies have shown collared, metadiaphyseal-filling (CMDF) cementless stems to have an equivalent periprosthetic fracture risk compared to cemented stems in primary total hip arthroplasty. This study assessed the associated risk of periprosthetic femoral fracture (PPFx) after hemiarthroplasty or total hip arthroplasty for femoral neck fracture comparing CMDF stems to cemented stems utilizing the American Joint Replacement Registry. METHODS: All femoral neck fractures in patients aged ≥ 65 years treated with arthroplasty in the American Joint Replacement Registry were linked to Centers for Medicare and Medicaid Services data from 2012 to 2022. Patients were included if they received a cementless CMDF stem or a cemented stem. Patient demographics, revision, or open reduction and internal fixation for PPFx, and revision for aseptic loosening or dislocation were recorded. Cause-specific Cox proportional hazard models were used to evaluate the association of stem design with all-cause revision, revision or open reduction and internal fixation due to PPFx, and revision for aseptic loosening or dislocation whereas adjusting for potential confounders. RESULTS: Compared to cemented stems, cementless CMDF stems were associated with a higher hazard of revision for PPFx (adjusted hazard ratio ] 1.7, 95% confidence interval: 1.1 to 2.7, P = 0.02). Analysis of secondary outcomes indicated that the two groups significantly differed in terms of revision for aseptic loosening, with cementless CMDF stems indicating a much lower event rate than cemented stems (adjusted hazard ratio: 0.5, 95% confidence interval: 0.2 to 0.9, P = 0.04). There were no significant differences found on the adjusted analyses of other secondary outcomes. CONCLUSIONS: Cementless CMDF stems were associated with an increased risk of revision for PPFx in the treatment of femoral neck fractures when compared to cemented stems. Although this finding supports the use of cemented stems for the treatment of femoral neck fractures, it should be assessed against an increased risk of aseptic loosening when compared to cementless CMDF stems.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Deckey DG, Kelly M, Zaniletti I, et al. Increased Risk of Periprosthetic Femoral Fracture With Cementless Collared Metadiaphyseal-Filling Stems Compared to Cemented Fixation for Femoral Neck Fractures in Patients Aged 65 Years or Older: An American Joint Replacement Registry Analysis. J Arthroplasty. 2026. PMID: 42617925. doi:10.1016/j.arth.2026.08.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。