文献速递关节外科

机器人全膝关节置换中切口内与切口外固定针的生物力学评估:尸体研究

Biomechanical Assessment of Pin-Site Fractures in Intra- Versus Extraincisional Pinning Technique for Robotic Total Knee Arthroplasty: A Cadaver Study.
PubMed 文献 · J Arthroplasty · 2026-07-16

该尸体研究将20对匹配的股骨和胫骨随机分配至切口内/干骺端或切口外/骨干固定针组,进行两阶段扭转负荷至失败测试。结果显示切口内股骨在100 Nm和300 Nm下扭转刚度均显著更高,整体骨折率显著更低(30% vs 80%,P=0.025),且无固定针部位骨折。

要点

  • 切口内固定针股骨在100 Nm和300 Nm下扭转刚度均显著高于切口外固定针。
  • 切口外标本整体骨折率80%,显著高于切口内标本的30%(P=0.025)。
  • 切口外标本中半数骨折涉及固定针部位,而切口内标本无固定针部位骨折。

核心结论

  • 切口内固定针置入在股骨和胫骨中均提供更高的扭转刚度和更低的骨折风险,尤其减少固定针部位失败。
  • 对于机器人全膝关节置换术后鼓励立即行走的患者,采用切口内固定针技术可能具有更好的安全性。

研究设计与样本

  • 尸体研究,20对匹配的股骨和胫骨随机分配至切口内/干骺端或切口外/骨干固定针组。
  • 标本骨水泥固定于夹具中,分两阶段进行扭转负荷至失败测试,先100 Nm后300 Nm,使用两台独立机器,连续监测扭矩、旋转和时间。

主要发现

  • 股骨平均扭转刚度在100 Nm下切口内为14.03 Nm/度,切口外为10.91 Nm/度(P=0.048);在300 Nm下切口内为14.46 Nm/度,切口外为10.91 Nm/度(P=0.048)。
  • 80%的切口外股骨在100 Nm扭转下骨折,而切口内股骨为20%(P=0.058)。
  • 胫骨平均扭转刚度切口内为8.24 Nm/度,切口外为6.35 Nm/度(P=0.2),差异无统计学意义。
  • 80%的切口外胫骨在100 Nm扭转下骨折,而切口内胫骨为40%。
  • 总体而言,80%的切口外标本骨折,显著高于切口内标本的30%(P=0.025);半数切口外标本骨折涉及固定针部位,切口内标本则无。

临床怎么用

  • 在机器人全膝关节置换中,可考虑采用切口内固定针置入技术以降低固定针部位骨折风险。
  • 对于术后鼓励立即行走的患者,切口内固定针可能提供更好的安全性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Robotic-assisted total knee arthroplasty offers enhanced precision, but introduces new challenges, including pin-site fracture risk from tracker fixation, with incidence rates of 0.06 to 4.8%. This study aimed to evaluate the effects of intra- versus extra-incisional pin placement on torsional rigidity to fracture failure in cadaver specimens. METHODS: Twenty matched cadaver femora and tibiae were randomized to undergo either intraincisional/metaphyseal or extraincisionalincisional/diaphyseal pinning. Specimens were cemented into jigs and subjected to torsional load-to-failure in two stages, first at 100 Nm, then 300 Nm, with two separate machines, with continuous monitoring of torque, rotation, and time. Fracture documentation and trendline analysis were performed to calculate torsional rigidity and coefficient of determination (R2). Fracture characteristics, including pin-site involvement, and fracture patterns were also recorded. RESULTS: Mean torsional rigidity was significantly higher for intraincisional femora at both 100 Nm (14.03 versus 10.91 Nm/degree, P = 0.048) and 300 Nm (14.46 versus 10.91 Nm/degree, P = 0.048). 80% of extraincisional femora fractured when subjected to 100 Nm of torsion, than 20% of intraincisional femora (P = 0.058). Mean torsional rigidity was 8.24 Nm/degree for intraincisional compared to 6.35 Nm/degree for extraincisional pinning for tibiae (P = 0.2), although this did not reach statistical significance. There were 80% of extraincisional tibiae that fractured when subjected to 100 Nm of torsion, compared to 40% of intraincisional tibiae. Overall, 80% of extraincisional specimens fractured, compared to a significantly lower 30% of intraincisional specimens (P = 0.025). There were half of the extraincisional specimens had fractures that involved the pin site(s), compared to none in intraincisional specimens. CONCLUSIONS: Intraincisional pin placement confers biomechanical advantages by having significantly higher torsional rigidity and reduced fracture risk, particularly of pin-site failures, in both femora and tibiae compared to extraincisional pin placements. Patients who are encouraged to ambulate immediately after robotic-assisted total knee arthroplasty operations may have a better safety profile by using intraincisional pinning techniques.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Liau ZQG, Shao-Rong Pang A, Kee Ern ET, et al. Biomechanical Assessment of Pin-Site Fractures in Intra- Versus Extraincisional Pinning Technique for Robotic Total Knee Arthroplasty: A Cadaver Study. J Arthroplasty. 2026. PMID: 42462879. doi:10.1016/j.arth.2026.07.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。