文献速递脊柱外科

三角钛植入物置于骨盆固定螺钉上方可降低多节段脊柱融合骨盆固定失败率:随机对照试验结果

Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion: Results of a Randomized Controlled Trial.
PubMed 文献 · J Bone Joint Surg Am · 2026-07-16

这项前瞻性、多中心、国际性、部分盲法随机对照试验纳入222例计划行多节段脊柱融合并采用S2AI骨盆固定的成人脊柱畸形患者,比较S2AI螺钉单独固定与S2AI+TTI固定的效果。结果显示,主要复合影像学终点在S2AI+TTI组发生率显著低于S2AI单独组(11.0% vs 20.4%,p=0.0415),且TTI放置未影响矢状面或冠状面脊柱序列参数。

要点

  • S2AI+TTI组主要复合影像学终点发生率为11.0%,低于S2AI单独组的20.4%(p=0.0415)。
  • TTI放置未影响矢状面或冠状面脊柱序列参数。
  • S2AI+TTI组全脊柱结构内断棒率略高(9.2% vs 5.3%),但差异无统计学意义(p=0.197)。

核心结论

  • 在行多节段脊柱融合并骨盆固定的患者中,将TTI置于S2AI螺钉上方可降低骨盆固定失败率。

研究设计与样本

  • 前瞻性、多中心、国际性、部分盲法随机对照试验,纳入222例计划行多节段脊柱融合并S2AI骨盆固定的成人脊柱畸形患者,随机分为S2AI单独组与S2AI+TTI组,计划随访24个月并行全脊柱X线及高分辨率CT独立判读。

主要发现

  • 主要复合影像学终点(S2AI螺钉断裂、透亮或拔出,或S1以远后路脊柱棒断裂)在S2AI+TTI组发生率为11.0%,低于S2AI单独组的20.4%(p=0.0415)。
  • 除 rod-screw dissociation 和S1以远断棒外,主要复合终点的其余组成部分在S2AI+TTI组发生率均更低。
  • TTI放置未影响矢状面或冠状面脊柱序列参数。
  • 全脊柱结构内断棒率在S2AI+TTI组为9.2%,S2AI单独组为5.3%,差异无统计学意义(p=0.197)。

临床怎么用

  • 对于计划行多节段脊柱融合并S2AI骨盆固定的成人脊柱畸形患者,可考虑在S2AI螺钉上方放置TTI以降低骨盆固定失败率。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study was to determine whether placement of triangular titanium implants (TTI) suprajacent to S2-alar-iliac (S2AI) pelvic fixation screws during multilevel spinal fusion surgery for adult spine deformity reduces pelvic fixation failures. METHODS: This prospective, multicenter, international, partially blinded randomized clinical trial included 222 patients scheduled for multilevel spinal fusion surgery with S2AI-based pelvic fixation. In the S2AI+TTI group, the mean age was 64.9 years, 54.1% were female, 3.7% were Hispanic, 5.5% were African American, and 1.8% were Asian. In the S2AI-alone group, the mean age was 64.8 years, 58.4% were female, 3.5% were Hispanic, 7.1% were African American, and 0.9% were Asian. Subjects were randomized to pelvic fixation using S2AI screws alone versus S2AI+TTI. Scheduled follow-up included a 24-month full-spine radiograph and a high-resolution computed tomography (CT) scan, which was independently interpreted. RESULTS: The primary composite radiographic end point (S2AI screw breakage, lucency, or pullout, or posterior spinal rod breakage distal to S1) occurred less frequently in the S2AI+TTI group than in the S2AI-alone group (11.0% versus 20.4%, p = 0.0415). Except for rod-screw dissociation and rod fracture distal to S1, all components of the primary composite end point occurred less frequently in the S2AI+TTI group. TTI placement did not affect sagittal or coronal spinal alignment parameters. The rate of rod fracture throughout the spinal construct was slightly higher in the S2AI+TTI group (9.2% versus 5.3%) but the difference was not significant (p = 0.197). CONCLUSIONS: Placement of TTI suprajacent to S2AI screws reduced the rate of pelvic fixation failure in patients undergoing multilevel spinal fusion with pelvic fixation. LEVEL OF EVIDENCE: Therapeutic Level I . See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Elder BD, Mundis G, Eastlack R, et al. Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion: Results of a Randomized Controlled Trial. J Bone Joint Surg Am. 2026. PMID: 42462059. doi:10.2106/JBJS.25.01415

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。