现代三锥度带领骨水泥柄全髋置换常见股骨距吸收

Calcar Resorption Is Commonly Seen With a Modern Triple-Tapered Collared Stem in Primary Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2026-04-15

该回顾性研究纳入182例接受直接前入路初次全髋关节置换并使用带领骨水泥柄的患者,平均随访2.3年。结果显示79.1%的患者出现股骨距吸收,表现为股骨距变薄和领部悬突增加,提示带领柄主要实现干骺端固定,领部对股骨距的载荷传递有限。

要点

  • 带领三锥度骨水泥柄全髋置换后股骨距吸收发生率达79.1%。
  • 股骨距平均宽度从术后初期5.9mm降至最终3.3mm。
  • 领部平均悬突从术后初期0.8mm增至最终3.4mm。

核心结论

  • 带领三锥度骨水泥柄主要实现干骺端固定,领部对股骨距的载荷传递有限。

研究设计与样本

  • 回顾性研究,纳入两家机构接受直接前入路初次全髋关节置换并使用带领骨水泥柄的患者182例,至少随访1年。

主要发现

  • 182例患者中144例(79.1%)出现股骨距吸收。
  • 股骨距吸收患者平均初始术后股骨距宽度5.9mm(SD 1.89),最终降至3.3mm(SD 1.84)。
  • 平均初始术后领部悬突0.8mm(SD 1.44),最终增至3.4mm(SD 2.6)。
  • 平均随访2.3年(范围1至6.2年),平均患者年龄64.2岁(范围18至83岁)。

临床怎么用

  • 全髋置换使用带领三锥度骨水泥柄时,应预期股骨距吸收的常见发生,需关注其临床意义尚不明确。

局限性

  • 股骨距变薄在带领柄背景下的临床意义尚未完全明确,需要未来研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The impacts of load transfer between the collar of modern triple-tapered collared hip stems and the femoral calcar remain unknown. The aim of this study was to determine whether the calcar remodels in response to stress loading or stress shielding in the setting of a well-fixed modern triple-tapered collared stem. METHODS: We conducted a retrospective review of patients from two institutions who underwent primary direct anterior total hip arthroplasty using collared cementless stems, with a minimum 1-year follow-up (n = 182). Statistical analyses were performed to determine associations between patients who demonstrated calcar resorption compared to those who did not. The mean follow-up was 2.3 years (range, one to 6.2), and the mean patient age was 64.2 years (range, 18 to 83). RESULTS: Calcar remodeling, more specifically resorption, occurred in 144 of 182 patients (79.1%). In patients who had resorption, the mean initial postoperative calcar width was 5.9 mm (standard deviation (SD), 1.89), decreasing to a mean final width of 3.3 mm (SD, 1.84). The mean initial postoperative collar overhang was 0.8 mm (SD, 1.44), increasing to a mean final collar overhang of 3.4 mm (SD = 2.6). CONCLUSIONS: Our results demonstrate that 79.1% of the patients who undergo primary total hip arthroplasty with a well-fixed triple-tapered collared cementless stem exhibit calcar resorption in the form of both calcar thinning and increasing collar overhang. These findings suggest that collared stems primarily achieve metaphyseal fixation. While the collar may decrease axial and rotational forces early on, the collar likely provides minimal load transfer to the calcar with stem ingrowth. Clinical implications of calcar thinning in the setting of a collared stem are yet to be fully understood. Future study of this finding is warranted.

来源:J Arthroplasty,查看原文。版权归原作者所有。