文献速递关节外科创伤骨科

575例旋转铰链式全膝关节置换术的无菌性松动率出乎意料地低

Surprisingly Low Rates of Aseptic Loosening in 575 Rotating-Hinge Total Knee Arthroplasties.
PubMed 文献 · J Bone Joint Surg Am · 2026-04-10

本研究回顾性分析了2002年至2021年间单中心575例旋转铰链式全膝关节置换术(RH-TKA),平均随访6年。结果显示10年无任何翻修生存率为64%,而无菌性松动翻修生存率高达90%;但因假体周围感染(PJI)二期再植入者翻修风险加倍。

要点

  • RH-TKA10年无无菌性松动翻修生存率达90%,为目前已发表最佳之一。
  • 因PJI二期再植入的RH-TKA任何翻修风险显著增加(HR=2)。
  • 未翻修且影像可评估的425膝中,股骨与胫骨假体松动征象分别占6%和8%。

核心结论

  • 当代RH-TKA的无菌性松动翻修率较低,10年无翻修生存率达90%。
  • 因PJI行二期再植入的RH-TKA失败风险显著高于无菌性病因者。

研究设计与样本

  • 单中心回顾性队列研究,纳入2002至2021年575例RH-TKA,平均年龄67岁,女性占58%,平均BMI 33 kg/m²,平均随访6年(范围2-19年)。

主要发现

  • 10年无任何翻修生存率为64%,5年为76%。
  • 翻修最常见原因为PJI(54%)和无菌性松动(20%)。
  • 无菌性病因者10年无任何翻修生存率79%,PJI再植入者为60%。
  • 10年无PJI翻修生存率总体81%,PJI再植入者仅73%。
  • 2年Knee Society Score从33分改善至69分(p<0.001)。

关键数值与推荐

  • 10年无无菌性松动翻修生存率为90%(5年为96%)。
  • PJI再植入是任何翻修的独立危险因素(HR=2,p<0.001)。
  • 影像学评估425膝中,股骨假体松动征象6%,胫骨假体8%。

临床怎么用

  • 因PJI行二期再植入者需告知翻修风险加倍,并加强随访监测。

局限性

  • 回顾性单中心设计,随访时间不均(2-19年),影像学评估仅限未翻修且有影像的425膝。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Contemporary rotating-hinge total knee arthroplasties (RH-TKAs) have shown reasonable short-term survivorship in smaller series, but concerns remain regarding risks of aseptic and septic failure. The purpose of this study was to assess outcomes of contemporary RH-TKAs in one of the largest series to date. METHODS: We retrospectively identified 575 RH-TKAs (60% used for aseptic etiologies and 40% used during reimplantation in 2-stage treatment of periprosthetic joint infection [PJI]) from 2002 to 2021 at a single institution. The mean age was 67 years, 58% were female, and the mean body mass index was 33 kg/m 2 . Sixty-five percent had Anderson Orthopaedic Research Institute (AORI) type-2B or 3 bone loss. Kaplan-Meier survivorship analyses were performed. The mean follow-up was 6 years (range, 2 to 19 years). RESULTS: Survivorship free from any revision was 76% at 5 years and 64% at 10 years. The most common revision indications were PJI (54%) and aseptic loosening (20%). RH-TKA used in the primary setting showed better survivorship compared with RH-TKA used during reimplantation after PJI (79% versus 60% at 10 years). Survivorship free from revision for aseptic loosening was 96% at 5 years and 90% at 10 years. Survivorship free from revision for PJI was 84% at 5 years and 81% at 10 years. Survivorship free from revision for PJI was even lower for RH-TKAs used during reimplantation, 74% at 5 years and 73% at 10 years. RH-TKA used during reimplantation in the treatment of PJI was associated with an increased risk of any revision (HR = 2, p < 0.001). Radiographic analysis of 425 knees that were not revised and had radiographs available for review showed that 6% of femoral components and 8% of tibial components had evidence of loosening at the time of final follow-up. The mean Knee Society Score improved from 33 to 69 at 2 years (p < 0.001). CONCLUSIONS: The 10-year survivorship free from aseptic loosening was 90% in this large series of RH-TKAs. This represents one of the best survivorships free from aseptic loosening published to date. Knees with prior PJI had markedly poorer survivorship than knees treated for aseptic etiologies, with double the risk of revision. LEVEL OF EVIDENCE: Therapeutic Level III . See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Terhune EB, Carstens MF, Fruth KM, et al. Surprisingly Low Rates of Aseptic Loosening in 575 Rotating-Hinge Total Knee Arthroplasties. J Bone Joint Surg Am. 2026. PMID: 41961950. doi:10.2106/JBJS.25.00837

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。