文献速递关节外科

双动髋臼与≥40mm大头假体在全髋关节置换术后的比较:一项基于注册队列的研究

Dual Mobility versus Jumbo Head (≥ 40 mm) Following Primary Total Hip Arthroplasty: A Registry-Based Cohort Study.
PubMed 文献 · J Arthroplasty · 2026-03-02

该研究基于美国注册队列比较了双动(DM)与≥40mm大头(JH)假体在初次全髋关节置换术后的脱位与翻修风险。共纳入9,532例(DM 1,716例,JH 7,816例),6年脱位发生率分别为2.4%和2.2%。DM在4.25年前脱位风险较低(HR 0.52),但之后风险较高(HR 8.26);全因、无菌及不稳定性翻修风险无显著差异。

要点

  • DM在术后4.25年前脱位风险低于JH,但后期脱位风险更高。
  • 两组全因、无菌及不稳定性翻修风险无显著差异。
  • 多数早期JH脱位可闭合复位成功,而早期DM脱位更常需翻修。

核心结论

  • DM与早期脱位减少相关,但可能后期脱位增多,后期差异小且随访有限。
  • 翻修风险(包括不稳定性翻修)在DM与JH间未检测到差异。

研究设计与样本

  • 基于美国注册数据的队列研究,纳入2010至2024年初次全髋关节置换术成人患者。
  • 共纳入9,532例髋(DM 1,716例,JH 7,816例),仅使用与两种假体均兼容的髋臼植入物以控制适应证和假体代次混杂。
  • 采用倾向评分加权Cox比例风险回归评估翻修风险。

主要发现

  • DM和JH的6年脱位发生率分别为2.4%和2.2%。
  • DM在4.25年前脱位风险较低(HR 0.52,95%CI 0.32-0.85),但之后风险较高(HR 8.26,95%CI 1.82-37.48)。
  • 全因翻修风险无差异(HR 1.22,95%CI 0.72-2.07)。
  • 无菌翻修风险无差异(HR 1.47,95%CI 0.73-2.93)。
  • 不稳定性翻修风险无差异(HR 1.71,95%CI 0.79-3.69)。

临床怎么用

  • DM可能减少早期脱位,但需关注后期脱位风险。
  • 两种大尺寸关节方案在翻修风险上相似,选择时可综合考虑复杂性和成本。
  • 早期JH脱位多可闭合复位,而DM脱位更常需翻修。

局限性

  • 后期随访有限,后期脱位差异需谨慎解读。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Instability remains a leading complication following primary total hip arthroplasty (THA). Large articulations enhance stability by increasing jump distance and range of motion. Dual mobility (DM) and ≥ 40 mm jumbo heads (JHs) are both large articulation options when compared to ≤ 36 mm heads, and debate exists between these two given differences in complexity, cost, and failure mechanism. METHODS: We conducted a united States registry-based cohort study to compare DM and JH. Adults who underwent primary THA between 2010 and 2024 were included. There were 9,532 TH As (1,716 DMs and 7,816 J Hs) included. To mitigate confounding by indication and by implant generation, only cases that used DM or JH with acetabular implants compatible with both options were included. The primary outcome was dislocation; secondary outcomes were all-cause revision, aseptic revision, and revision for instability. Crude incidence was calculated at six years. Propensity score-weighted Cox proportional hazards regressions were used to evaluate revision risk. RESULTS: Dislocation incidence was 2.4 and 2.2% for DMs and JHs, respectively. We observed a time-based difference in dislocation risk: DM was associated with a lower risk before 4.25 years (hazard ratio (HR): 0.52, 95% confidence interval (CI): 0.32 to 0.85), but a higher risk after 4.25 years (HR: 8.26, 95% CI: 1.82 to 37.48). There were no differences found in all-cause (HR: 1.22, 95% CI: 0.72 to 2.07), aseptic (HR: 1.47, 95% CI: 0.73 to 2.93), or instability-related revision risk (HR: 1.71, 95% CI: 0.79 to 3.69). CONCLUSIONS: Use of a DM was associated with fewer early dislocations, but possibly more late dislocations, although late differences were small and later follow-up was limited. There was also no detectable difference in revisions, including for instability. While DM's marginal size advantage may avoid a small number of early closed dislocations, current large articulation options may be sufficiently similar to mitigate major revision differences. Most early JH dislocations were successfully treated closed, and conversely, the fewer early DM dislocations more often required revision.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Chen F, Prentice HA, Fasig BH, et al. Dual Mobility versus Jumbo Head (≥ 40 mm) Following Primary Total Hip Arthroplasty: A Registry-Based Cohort Study. J Arthroplasty. 2026. PMID: 41780831. doi:10.1016/j.arth.2026.02.038

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。