双动髋臼组件在全髋关节置换术中的长期生存率:25545例回顾性分析
基于法国SOFCOT/RENACOT登记系统58314例初次全髋关节置换术的回顾性研究,比较双动组件(DMC,25545例)与传统组件(CC)的长期生存率。调整混杂因素后,DMC并未显著增加翻修风险(HR 0.83,95%CI 0.66-1.04,p=0.118)。DMC组假体周围骨折翻修比例更高(34.5% vs 15.6%,p<0.001),但脱位翻修更少,脱位或骨折合计翻修比例更低(39.4% vs 51.5%)。
要点
- DMC组翻修率1.1%(284/25545),调整后翻修风险与CC无显著差异(HR 0.83,p=0.118)。
- DMC组假体周围骨折翻修比例显著高于CC组(34.5% vs 15.6%,p<0.001)。
- DMC组脱位翻修更少,脱位或骨折合计翻修比例低于CC组(39.4% vs 51.5%)。
核心结论
- DMC在初次全髋关节置换中提供良好的长期生存率,并有效降低脱位风险。
- 尽管DMC组假体周围骨折发生率更高,总体结果支持其安全有效使用,仍需持续监测该风险。
研究设计与样本
- 基于法国国家SOFCOT/RENACOT登记系统的回顾性研究,分析2006年1月至2023年12月58314例初次全髋关节置换术,其中DMC 25545例(46%)。
- 主要结局为任何原因翻修,采用Kaplan-Meier生存曲线和Cox比例风险模型比较,并调整年龄、性别、主要诊断和固定方式。
主要发现
- DMC组记录284例翻修(1.1%)。
- 调整后Cox回归显示DMC与CC相比翻修风险无显著增加(HR 0.83,95%CI 0.66-1.04,p=0.118)。
- DMC组假体周围骨折是翻修首要原因(98/284,34.5%),显著高于CC组(77/493,15.6%,p<0.001)。
- DMC组脱位相关翻修更少见,脱位或骨折合计翻修比例DMC组为39.4%(112/284),CC组为51.5%(254/493)。
- 多变量Cox回归未发现假体失败的危险因素。
临床怎么用
- 对于初次全髋关节置换患者,DMC可作为一种有效降低脱位风险的组件选择。
- 使用DMC时需关注假体周围骨折风险,尤其在高风险患者中应加强监测。
局限性
- 本研究为基于登记系统的回顾性系列,可能存在未测量混杂因素。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: This registry-based study aimed to evaluate the long-term survivorship of dual-mobility components (DMCs) compared to conventional components (CCs) in primary total hip arthroplasty (THA), and to assess potential adverse effects, particularly in younger patients (aged < 75 years). METHODS: Data from 58,314 primary THAs recorded between January 2006 and December 2023 in the French national SOFCOT/RENACOT registry were analyzed, including 25,545 DMCs (46%). The primary outcome was revision for any cause. Kaplan-Meier survival curves and Cox proportional hazards models were used to compare implant survivorship, adjusting for age, sex, primary diagnosis, and fixation method. RESULTS: Among the 25,545 primary THAs performed with a DMC, 284 revisions (1.1%) were recorded. In adjusted Cox regression analysis, DMC use was not significantly associated with an increased risk of revision compared to CCs (hazard ratio 0.83 (95% CI 0.66 to 1.04); p = 0.118). Periprosthetic fracture was the leading cause of revision in the DMC group (98/284, 34.5%), occurring significantly more frequently than in the CC group (77/493, 15.6%; p < 0.001), while dislocation-related revisions were less common with DMCs. Overall, the combined proportion of revisions due to dislocation or fracture was lower in the DMC group (112/284, 39.4%) than in the CC group (254/493, 51.5%). No risk factors for implant failure were identified in multivariable Cox regression analysis. CONCLUSION: DMCs provide excellent long-term survivorship and effectively reduce the risk of dislocation in primary THA. Although a higher incidence of periprosthetic fractures was observed with DMCs, the overall outcomes support their safe and effective use. Nevertheless, ongoing surveillance remains important to monitor these risks.
引用
- Bouché PA, Brand C, Putman S, et al. Total hip arthroplasty with dual-mobility components : a retrospective series of 25,545 primary cases. Bone Joint J. 2026. PMID: 41763256. doi:10.1302/0301-620X.108B3.BJJ-2025-0766.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。