棒直径≤4.5 mm与断棒风险增加2.46倍相关,较轻患者螺钉拔出、内固定突出、皮肤破溃及深部感染风险更高。

要点

  • 断棒多发生于畸形顶点附近两个椎体范围内,螺钉拔出多位于内固定上端或下端椎附近。
  • 棒直径≤4.5 mm较>4.5 mm断棒风险增加2.46倍(P=0.039)。
  • 较轻患者螺钉拔出风险增加2.0倍、内固定突出3.1倍、皮肤破溃7.7倍、深部感染3.6倍(均P<0.05)。

核心结论

  • 生长引导手术内固定失败以断棒和螺钉拔出为主,断棒集中于畸形顶点,螺钉拔出集中于内固定两端。
  • 使用直径>4.5 mm的棒、确保牢固的近远端固定并最大化肌肉包覆可能减少并发症及再手术。

研究设计与样本

  • 回顾性、多中心队列研究,分析多中心早发性脊柱侧凸数据库中接受生长引导手术的患者。

主要发现

  • 共发生173例次内固定并发症:断棒55例次(32%)、螺钉拔出46例次(27%)、内固定突出33例次(19%)、植入物表面皮肤破溃7例次(4%)、患者生长超出棒长度32例次(18%)。
  • 棒直径≤4.5 mm较>4.5 mm断棒风险增加2.46倍(P=0.039)。
  • 较轻患者较重的患者螺钉拔出风险增加2.0倍、内固定突出风险增加3.1倍、植入物表面皮肤破溃风险增加7.7倍、深部伤口感染风险增加3.6倍(均P<0.05)。
  • 断棒更多发生于距畸形顶点两个椎体范围内,螺钉拔出更多位于内固定上端或最下端椎附近。

临床怎么用

  • 选用直径>4.5 mm的棒可能降低断棒风险。
  • 确保近端和远端牢固固定,并尽量保留肌肉包覆,可能减少内固定突出、皮肤破溃及感染。
  • 对较轻(多较年轻)患者应加强感染、内固定突出和皮肤完整性监测。

局限性

  • 研究为回顾性设计,证据等级为Level III。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Retrospective, multicenter cohort study. SUMMARY OF BACKGROUND DATA: Growth guidance surgery (GGS) has lower reoperation rates and fewer episodes of care than other growth-sparing methods but is associated with instrumentation complications. This study aimed to characterize factors associated with implant complications in GGS. METHODS: A multicenter early-onset scoliosis database was analyzed for patients who underwent GGS. Radiographs, complication reports, and reoperation notes were evaluated for instances of rod breakage, screw pullout, instrumentation prominence, skin breakdown over implants, patients outgrowing implants, and deep wound infection. Descriptive statistics, χ 2 test, and Cox proportional-hazards models were utilized to detect differences in complications. RESULTS: One hundred eighteen patients [7±2 y old; 69 (58%) female] were included. Mean follow-up was 5 (±3) years. The 173 instances of instrumentation complications comprised 55 (32%) broken rods, 46 (27%) screw pullouts, 33 (19%) prominences, seven (4%) skin breakdown over implant, and 32 (18%) outgrowing the rods. There was a 2.46-fold ( P =0.039) increased risk of rod breakage for rod diameter ≤4.5 versus >4.5 mm. For lighter patients, odds were greater for screw pullout by 2.0-fold, for prominence by 3.1-fold, for skin breakdown over implant by 7.7-fold, and for deep wound infection by 3.6-fold compared with heavier patients (all P <0.05). Broken rods more often occurred within two vertebrae of the apex of deformity; screw pullouts near the upper or lowest instrumented vertebrae. CONCLUSIONS: Most instrumentation failures were broken rods and screw pullouts. Breakage was largely confined to the deformity apex and was mitigated by rods >4.5 mm; pullouts clustered at construct ends. Lighter ( i.e. mostly younger) patients faced higher risks of infection, prominence, and skin compromise. Using larger-diameter rods, ensuring robust proximal-distal fixation, and maximizing the muscle envelope may meaningfully reduce these complications and, thus, reoperations. LEVEL OF EVIDENCE: Level III.