高压蒸汽灭菌可有效灭菌骨科手术托盘中受污染的残留骨水泥

Autoclaving Effectively Sterilizes Contaminated Retained Cement in Orthopaedic Surgical Trays.
PubMed 文献 · J Bone Joint Surg Am · 2026-01-23

本研究在聚甲基丙烯酸甲酯(PMMA)试片(光滑面或深裂隙,每种条件n=5)及螺纹螺钉(插入聚合中的PMMA,每种条件n=5)上培养成熟的金黄色葡萄球菌(MSSA)和大肠埃希菌生物膜,经3种临床高压蒸汽灭菌方案处理后检测菌落形成单位(CFU)和三磷酸腺苷(ATP)水平。结果:未灭菌PMMA表面污染严重,MSSA为4.1 ± 2.3 × 10^6 CFUs/mL,大肠埃希菌为2.4 ± 1.2 × 10^6 CFUs/mL;灭菌后所有PMMA表面、裂隙和螺钉的CFUs/mL均<10、RLU均<5,扫描电镜显示生物膜结构完整但无活菌。结论:临床高压蒸汽灭菌方案可有效灭菌被革兰阳性或革兰阴性生物膜污染的PMMA及PMMA覆盖器械。

要点

  • 未灭菌PMMA表面MSSA污染量为4.1 ± 2.3 × 10^6 CFUs/mL,大肠埃希菌为2.4 ± 1.2 × 10^6 CFUs/mL。
  • 高压蒸汽灭菌后所有PMMA表面、裂隙和螺钉的CFUs/mL均<10、RLU均<5。
  • 扫描电镜显示生物膜结构完整,但未检出活菌。

核心结论

  • 被革兰阳性或革兰阴性生物膜污染的PMMA及PMMA覆盖器械经临床高压蒸汽灭菌方案可有效灭菌。

研究设计与样本

  • 体外研究:在PMMA试片(光滑面或深裂隙,每种条件n=5)及螺纹螺钉(插入聚合中的PMMA,每种条件n=5)上培养成熟的MSSA和大肠埃希菌生物膜,并采用3种高压蒸汽灭菌方案处理。

主要发现

  • 未灭菌PMMA表面MSSA污染量为4.1 ± 2.3 × 10^6 CFUs/mL,大肠埃希菌为2.4 ± 1.2 × 10^6 CFUs/mL;ATP水平分别为3,230 ± 1,297 RLU和1,006 ± 602 RLU。
  • 裂隙和螺钉同样严重污染。
  • 高压蒸汽灭菌后所有PMMA表面、裂隙和螺钉的CFUs/mL均<10、RLU均<5(低于清洁卫生阈值)。
  • 扫描电镜显示生物膜结构完整,但无活菌。

临床怎么用

  • 经临床高压蒸汽灭菌方案处理的残留PMMA及PMMA覆盖手术器械不携带活菌。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Retained polymethylmethacrylate (PMMA) in surgical instrument trays and on instruments is becoming more common and is a disquieting situation for the arthroplasty operating room team. Although retained debris may be sterile after autoclaving, previous studies have not specifically investigated bacterial viability on retained PMMA. This study determined whether heavily contaminated PMMA surfaces and instruments covered in PMMA could be effectively sterilized by clinically utilized autoclaving protocols. METHODS: Mature methicillin-susceptible Staphylococcus aureus (MSSA) and Escherichia coli biofilms were grown on PMMA coupons with a smooth surface (n = 5 per condition) or with deep crevices (n = 5 per condition) and on threaded screws, which were inserted into polymerizing PMMA (n = 5 per condition). PMMA coupons and PMMA-covered screws were subjected to 3 autoclave sterilization protocols. Colony-forming unit (CFU) counts and adenosine triphosphate (ATP) levels were measured to detect viable bacteria. Scanning electron microscope (SEM) images were obtained to determine whether biofilm was present. RESULTS: Non-autoclaved PMMA surfaces were heavily contaminated, with values of 4.1 ± 2.3 × 10 6 CFUs/mL for S . aureus and 2.4 ± 1.2 × 10 6 CFUs/mL for E . coli. ATP levels of 3,230 ± 1,297 relative light units (RLU) for S . aureus and 1,006 ± 602 RLU for E . coli were observed. Crevices and screws were also heavily contaminated. Following autoclaving, CFUs/mL counts of <10 and RLU levels of <5 (beneath the hygienic threshold for cleanliness) were recorded on all PMMA surfaces, crevices, and screws. SEM confirmed that the biofilm structures remained intact, but they did not contain viable bacteria. CONCLUSIONS: PMMA and PMMA-covered instruments contaminated with gram-positive or gram-negative biofilms are effectively sterilized using clinical autoclaving protocols. CLINICAL RELEVANCE: Retained PMMA and PMMA-covered surgical instruments that undergo a clinical autoclaving protocol do not carry viable bacteria.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。