增强现实导航与徒手椎弓根螺钉置入的准确性比较
这项回顾性队列研究分析了202例患者共1109枚腰骶椎椎弓根螺钉,比较增强现实(AR)导航与徒手(FH)置入的准确性。AR组临床可接受螺钉比例显著高于FH组(99.3% vs 96.4%,P=0.013),不可接受突破率显著更低(0.7% vs 3.6%,P=0.013)。作者认为AR可能提供更一致的置入准确性,需进一步前瞻性研究确定长期临床意义。
要点
- AR组临床可接受螺钉(GRS A或B级)比例99.3%,显著高于FH组96.4%(P=0.013)。
- AR组临床不可接受突破(GRS C-E级)率0.7%,显著低于FH组3.6%(P=0.013)。
- AR组突破最多见于L4且以偏外侧为主(n=24),FH组最多见于L5且以偏内侧为主(n=22)。
核心结论
- AR导航与传统FH技术相比,椎弓根螺钉置入准确性显著更高。
- 作为迄今直接比较这两种方法的最大样本研究,本分析支持AR可能提供更一致螺钉置入准确性的观点。
- 需进一步前瞻性研究确定其长期临床意义。
研究设计与样本
- 回顾性队列研究,纳入所有接受L1至S1椎弓根螺钉置入(AR导航或传统FH技术)且有术后CT影像可用于螺钉准确性评估的患者。
- 共分析202例患者、1109枚椎弓根螺钉,其中AR组548枚、FH组561枚。
- 采用基于CT的Gertzbein-Robbins(GRS)分级评估准确性,并比较两组人口学和手术特征。
主要发现
- 临床可接受螺钉(GRS A或B级)比例AR组99.3% vs FH组96.4%(P=0.013),AR组显著更高。
- 临床不可接受突破(GRS C-E级)率AR组0.7% vs FH组3.6%(P=0.013),AR组显著更低。
- 按节段的方向汇总显示,AR组突破最多见于L4、以偏外侧为主(n=24),FH组突破最多见于L5、以偏内侧为主(n=22)。
- 两组人口学和手术特征无显著差异。
关键数值与推荐
- 螺钉准确性按Gertzbein-Robbins(GRS)分级判定,A或B级为临床可接受,C-E级为临床不可接受突破。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文指出需进一步前瞻性研究确定其长期临床意义。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: Retrospective cohort study. OBJECTIVE: To compare the accuracy of augmented reality (AR)-guided and freehand (FH) pedicle screw placement in the lumbosacral spine using computed tomography (CT)-based Gertzbein-Robbins (GRS) grading, and to assess demographic and surgical factors associated with accuracy. BACKGROUND: Accurate pedicle screw placement is essential for safe and effective spinal fusion. AR navigation has emerged as an alternative to conventional FH techniques, with potential advantages in accuracy, efficiency, and ergonomics. MATERIALS AND METHODS: All patients who underwent pedicle screws placement from L1 to S1 using either AR navigation or conventional FH technique, with available postoperative CT imaging for screw accuracy assessment, were analyzed. Demographic and surgical characteristics were compared between both cohorts. RESULTS: A total of 202 patients were analyzed, comprising 1109 pedicle screws placed (AR: n = 548; FH: n = 561). The proportion of clinically acceptable screws (GRS grade A or B) was significantly higher in the AR group compared with the FH group (99.3% vs . 96.4%; P = 0.013). The rate of clinically unacceptable breaches (GRS grade C-E) was significantly lower in the AR group than in the FH group (0.7% vs . 3.6%; P = 0.013). Level-specific directional summary showed that the highest number of breaches in the AR group occurred at L4, predominantly lateral (n = 24), while in the FH group, breaches were most common at L5 and primarily medial (n = 22). Demographic and surgical characteristics demonstrated no significant differences between the cohorts. CONCLUSIONS: AR-guided navigation was associated with significantly higher pedicle screw placement accuracy compared with conventional FH technique. As the largest study to date directly comparing these approaches, this analysis reinforces the notion that AR may offer more consistent accuracy in screw placement. Further prospective studies are needed to determine its long-term clinical implications.
引用
- Loggia G, Alakmeh A, Avrumova F, et al. Accuracy of Augmented Reality Navigation Versus Freehand Pedicle Screw Placement. Spine (Phila Pa 1976). 2025. PMID: 41396203. doi:10.1097/BRS.0000000000005594
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。