黏弹性椎间盘与两种关节型全椎间盘置换器磨损颗粒的体外对比分析
AxioMed装置平均磨损率为1.7 mg/MC,低于prodisc L的5.7 mg/MC,且颗粒质量平均直径49 μm大于CHARITÉ的0.2 μm和prodisc L的0.4 μm;
要点
- AxioMed装置平均磨损率1.7 mg/MC,prodisc L为5.7 mg/MC。
- AxioMed颗粒质量平均直径49 μm,CHARITÉ为0.2 μm,prodisc L为0.4 μm。
核心结论
- AxioMed一体式黏弹性全椎间盘置换装置与关节型全椎间盘置换器相比,磨损率更低、颗粒更大且生物反应性更低,提示骨溶解风险降低和植入物寿命延长。
- 本研究聚焦颗粒尺寸,需进一步研究以表征颗粒成分和颗粒负荷。
研究设计与样本
- CHARITÉ和prodisc L的对比数据来自美国食品药品监督管理局(FDA)安全性与有效性数据摘要。
主要发现
- AxioMed装置平均磨损率为1.7 mg/MC,prodisc L为5.7 mg/MC。
- AxioMed装置数量平均颗粒直径为1.9 μm,质量平均颗粒直径为49 μm,大于文献报告的CHARITÉ(0.2 μm)和prodisc L(0.4 μm)装置。
关键数值与推荐
- 大于1.0 μm的颗粒较少可能诱导炎症反应。
临床怎么用
- 作者认为该一体式黏弹性全椎间盘置换装置可能为保留运动的腰椎手术提供更安全、更耐用的替代选择。
- 作者指出需进一步开展临床和取回研究。
局限性
- 本研究聚焦颗粒尺寸,需进一步研究以表征颗粒成分和颗粒负荷。
- 需进一步开展临床和取回研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Wear debris is a known contributor to orthopaedic implant failure, particularly in joint arthroplasty. The wear characteristics of spinal total disc replacement (TDR) remain under-investigated. Spinal TDRs have been shown to produce wear particles that elicit strong inflammatory reactions. Submicron debris, in particular, is associated with osteolysis and implant loosening. Viscoelastic TDR (VTDR) devices have emerged to address these risks. METHODS: Five AxioMed Freedom Lumbar Disc (FLD) devices underwent 30 million cycles (10 million each in flexion-extension, lateral bending, and axial rotation) of wear testing in phosphate-buffered saline solution at 37°C using an MTS servohydraulic system. Wear fluid samples were collected every 5 million cycles and analyzed using scanning electron microscopy and laser diffraction. A 30-million device cycle count simulates 240 years of lumbar bending. Wear rates were calculated in milligrams per million cycles (mg/MC). Comparative data for CHARITÉ (DePuy Synthes) and prodisc L (Centinel Spine) discs were obtained from the United States Food and Drug Administration (FDA) Summary of Safety and Effectiveness Data. RESULTS: The AxioMed device showed a mean wear rate of 1.7 mg/MC, in comparison to 5.7 mg/MC for the prodisc L. The number-average particle diameter was 1.9 μm, with a mass-average particle diameter of 49 μm, which was notably larger than those reportedly produced by the CHARITÉ (0.2 μm) and prodisc L (0.4 μm) devices, which is promising because larger particles (>1.0 μm) are less likely to induce inflammatory responses. No mechanical failures were observed during the 30 million cycles. CONCLUSIONS: The AxioMed 1-piece VTDR device demonstrated a lower wear rate and larger, less biologically reactive, particles compared with articulating TDRs, suggesting a reduced risk of osteolysis and longer implant lifespan. No mechanical failures were observed, even after each 10-million-cycle interval, which simulates approximately 80 years of lumbar-bending motions. This study focused on particle size; further work is warranted to characterize composition and particle burden. CLINICAL RELEVANCE: This 1-piece VTDR may offer a safer and more durable alternative for motion-preserving lumbar spine surgery. Further clinical and retrieval studies are warranted.
引用
- Chin KR, Chebrolu S, Sung RD, et al. Comparative in Vitro Analysis of Wear Particles Generated by a Viscoelastic Disc Versus 2 Articulating Total Disc Replacements. J Bone Joint Surg Am. 2025. PMID: 41284766. doi:10.2106/JBJS.25.00594