滑液中快速髓过氧化物酶试验对慢性假体周围关节感染的诊断价值
研究纳入23例假体周围关节感染(PJI)和23例无菌性失败患者,比较滑液髓过氧化物酶(MPO)与α-防御素ELISA的诊断准确性,并评估MPO即时检验(POCT)。MPO与α-防御素ELISA强相关,最佳截断值下两者敏感性0.957、特异性0.957;MPO POCT 10分钟出结果,敏感性1.00、特异性0.913。
要点
- MPO ELISA与α-防御素ELISA强相关,最佳截断值下敏感性和特异性均为0.957。
- MPO POCT仅需10分钟,敏感性1.00、特异性0.913。
- MPO POCT有望成为诊断PJI的快速工具,并可作为α-防御素试验的替代方案。
核心结论
- MPO ELISA对PJI具有高诊断准确性,与α-防御素试验相当。
- MPO POCT同样表现出高诊断准确性,是诊断PJI的有前景的快速工具,并可能作为α-防御素试验的替代方案。
研究设计与样本
- 根据第2届国际共识会议(ICM)PJI诊断标准,将全关节置换术后疑似感染患者分为PJI组或无菌性失败组。
- 研究纳入23例PJI患者和23例无菌性失败患者,均为日本人。
- 采用酶联免疫吸附试验(ELISA)定量检测术中采集的滑液中MPO和α-防御素水平,并比较其诊断准确性;同时进行MPO POCT并评估其诊断准确性。
- 证据等级为诊断Level II。
主要发现
- MPO与α-防御素ELISA表现出强相关性。
- 在最佳截断值时,MPO ELISA和α-防御素ELISA的敏感性均为0.957,特异性均为0.957。
- MPO POCT在仅10分钟内产生结果,敏感性为1.00,特异性为0.913。
关键数值与推荐
- PJI与无菌性失败的分类依据第2届ICM PJI诊断标准。
- MPO ELISA在最佳截断值时敏感性0.957、特异性0.957。
- α-防御素ELISA在最佳截断值时敏感性0.957、特异性0.957。
- MPO POCT敏感性1.00、特异性0.913,出结果时间10分钟。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Numerous biomarkers have been identified for the diagnosis of periprosthetic joint infection (PJI), but no single biomarker has been proven to have definitive accuracy. Alpha-defensin is an excellent biomarker that is included as a minor criterion in the 2nd International Consensus Meeting (ICM) PJI diagnostic criteria; however, its high cost limits its accessibility at many general medical facilities. Consequently, alternative biomarkers for PJI diagnosis are under investigation. Myeloperoxidase (MPO), an enzyme with microbicidal properties through the catalysis of hypochlorous acid production, has emerged as a promising alternative. MPO has previously been reported as an effective biomarker for PJI. In the present study, we conducted a rapid MPO test as a point-of-care test (POCT) and evaluated its diagnostic utility. METHODS: Patients with a suspected infection after total joint arthroplasty were categorized into the PJI or aseptic failure groups according to the 2nd ICM PJI diagnostic criteria. MPO and alpha-defensin levels in synovial fluid that was collected during surgery were quantified using an enzyme-linked immunosorbent assay (ELISA), and their diagnostic accuracy was compared. Additionally, an MPO POCT was conducted, and its diagnostic accuracy was assessed. RESULTS: The investigation included 23 patients with PJI and 23 with aseptic failure. All were Japanese. MPO and alpha-defensin ELISAs exhibited a strong correlation, and, at the optimal cutoff, both tests demonstrated a sensitivity of 0.957 and a specificity of 0.957. The MPO POCT produced results in just 10 minutes and achieved a sensitivity of 1.00 and a specificity of 0.913. CONCLUSIONS: The MPO ELISA demonstrated high diagnostic accuracy for PJI, equivalent to that of the alpha-defensin test. Furthermore, the MPO POCT showed similarly high diagnostic accuracy. The MPO POCT is a promising and rapid tool for diagnosing PJI and could serve as an alternative to the alpha-defensin test. LEVEL OF EVIDENCE: Diagnostic Level II. See Instructions for Authors for a complete description of levels of evidence.